Agilent Technologies E8257D/67D manual E8267D Option, Option 532

Page 35

Operation Verification

Checking the Output Power

Table 3-5

Leveled Output Power Limits, E8257D cont.

Option 550

Frequency

Amplitude

 

Limits

 

 

 

 

 

Option

 

dBm

dB

 

 

 

 

 

60 GHz

None

 

5

±1.2

 

 

 

 

 

 

1EA

 

14

±1.2

 

 

 

 

 

 

1E1

 

3

±1.2

 

 

 

 

 

 

1E1 & 1EA

 

10

±1.2

 

 

 

 

 

67 GHz

None

 

5

±1.2

 

 

 

 

 

 

1EA

 

12

±1.2

 

 

 

 

 

 

1E1

 

3

±1.2

 

 

 

 

 

 

1E1 & 1EA

 

10

±1.2

 

 

 

 

 

Option 567

Frequency

Amplitude

Limits

 

 

 

 

 

Option

dBm

dB

 

 

 

 

60 GHz

None

5

±1.2

 

 

 

 

 

1EA

14

±1.2

 

 

 

 

 

1E1

3

±1.2

 

 

 

 

 

1E1 & 1EA

10

±1.2

 

 

 

 

67 GHz

None

5

±1.2

 

 

 

 

 

1EA

12

±1.2

 

 

 

 

 

1E1

3

±1.2

 

 

 

 

 

1E1 & 1EA

10

±1.2

 

 

 

 

Table 3-6

Leveled Output Power Limits, E8267D

E8267D

Option 520

Frequency

Amplitude

 

Limits

 

 

 

 

 

Option

 

dBm

dB

 

 

 

 

 

250 kHz

None

 

15

±0.6

 

 

 

 

 

3.2 GHz

None

 

15

±0.8

 

 

 

 

 

20 GHz

None

 

18

±0.8

 

 

 

 

 

Option 532 & 544

Frequency

Amplitude

Limits

 

 

 

 

 

Option

dBm

dB

 

 

 

 

250 kHz

None

12

±1.0

 

 

 

 

3.2 GHz

None

12

±1.0

 

 

 

 

20 GHz

None

14

±1.0

 

 

 

 

40 GHz

None

12

±1.0

 

 

 

 

44 GHz

None

10

±0.9

 

 

 

 

NOTE

Limit values are due to power meter uncertainty.

 

 

Chapter 3

31

Image 35 Contents
Installation Guide Questions or Comments about our Documentation? Contents Regulatory Information Instrument Markings Safety InformationSafety Information General Safety ConsiderationsGetting Started Checking the Shipment Chapter Meeting Electrical and Environmental RequirementsEnvironment VentilationConnecting the AC Power Cord Line SettingsAC Power Cord Localization Screen Saver Configuring the DisplayContrast and Brightness Inverse VideoUpdating the Display During Remote Operation LAN Interface Configuration Configuring for Remote ControlGpib Interface Configuration Set LAN Config Manual Dhcp to Dhcp Dhcp ConfigurationRS-232 Interface Configuration Front Handle and Rack Mount Flange Kits Ordering AccessoriesFront Handles and Rack Mount Flanges E8257D/67D PSG DocumentationOrdering Accessories Cleaning Suggestions Proper Usage and CleaningContacting Agilent Sales and Service Offices Returning a Signal Generator to Agilent Technologies Returning a Signal Generator to Agilent Technologies Operation Verification Performing a Self-Test Self-Test Failure Maximum Leveled Power Check Failure Checking the Maximum Leveled PowerAmplitude Limits Checking the Output Power Recommended Agilent Option 540Recommended Equipment for the E8267D 532 544 Checking the Output Power Amplitude Limits Option DBm E8257D OptionAmplitude Limits Option 1E1 ±0.9 1E1 & 1EA ±1.0 50 GHz None ±1.2 Option 532 E8267D OptionProblems with Output Power Regulatory Information Compliance with German Noise Requirements Statement of ComplianceAssistance CertificationCompliance with Canadian EMC Requirements Compliance with Canadian EMC Requirements Index IndexIndex