Operation Verification

Checking the Output Power

Perform the following procedure to check the signal generator output power:

1. Zero and calibrate the power sensor to the power meter as shown:

2. Connect the equipment as shown:

3.Press Preset.

4.Press RF On/Off. The RF On annunciator is now displayed.

5.Press Mod On/Off. The Mod Off annunciator is now displayed.

6.Press Frequency and enter the first frequency value listed in Table 3-1.

7.Press Amplitude and enter the amplitude value for your option and frequency value.

8.Measure the power level.

9.Repeat step 6 through 8 and measure the power levels for each remaining frequency listed in Table 3-4thru Table 3-6. Record any power levels not within the acceptable limits.

28

Chapter 3

Image 32 Contents
Installation Guide Questions or Comments about our Documentation? Contents Regulatory Information Safety Information Instrument MarkingsGeneral Safety Considerations Safety InformationGetting Started Checking the Shipment Meeting Electrical and Environmental Requirements EnvironmentVentilation ChapterLine Settings Connecting the AC Power CordAC Power Cord Localization Configuring the Display Contrast and BrightnessInverse Video Screen SaverUpdating the Display During Remote Operation LAN Interface Configuration Configuring for Remote ControlGpib Interface Configuration Dhcp Configuration Set LAN Config Manual Dhcp to DhcpRS-232 Interface Configuration Ordering Accessories Front Handles and Rack Mount FlangesE8257D/67D PSG Documentation Front Handle and Rack Mount Flange KitsOrdering Accessories Proper Usage and Cleaning Cleaning SuggestionsContacting Agilent Sales and Service Offices Returning a Signal Generator to Agilent Technologies Returning a Signal Generator to Agilent Technologies Operation Verification Performing a Self-Test Self-Test Failure Checking the Maximum Leveled Power Maximum Leveled Power Check FailureAmplitude Limits Checking the Output Power Recommended Agilent Option 540Recommended Equipment for the E8267D 532 544 Checking the Output Power Amplitude Limits Option DBm E8257D OptionAmplitude Limits Option 1E1 ±0.9 1E1 & 1EA ±1.0 50 GHz None ±1.2 E8267D Option Option 532Problems with Output Power Regulatory Information Statement of Compliance AssistanceCertification Compliance with German Noise RequirementsCompliance with Canadian EMC Requirements Compliance with Canadian EMC Requirements Index IndexIndex