Epson 214D-1 user manual Trouble Shooting

Page 53

Trouble Shooting 45

8. Trouble Shooting

Batcher does not reset

The Signal Timeout has been set to an excessively long period and has not timed out at the end of the last batch.

Batch will not start or relay 1 will not close

Ensure that the instrument has not timed out as controlled by the Signal Timeout and that a Flow Alarm condition does not prevail. Pressing the Stop switch will cancel this condition. Check for a fault on the flow input before restarting.

Batcher stops during a batch (prior to batch end)

This could be due to the Signal Timeout having timed out. Check for a fault in the system. Ensure that the Signal Timeout period is significantly longer than the period between flowmeter pulses at the minimum flowrate.

No display

Check for power to the instrument.

All 88888888 displayed

The Batcher displays all eights on power-up for 4 seconds as a display test. If all eights continue to display after this period, this is symptomatic of the power supply voltage being low. Check the power input voltage.

Not counting.

If the Batcher does not count with the flowmeter connected and flow passing through it, first check the connections and then ensure the DIP switches on the rear of the instrument are set correctly for the attached flowmeter.

It is possible to manually test the input circuit of the Batcher by setting the input configuration for a Reed Switch and pulsing across the signal (+) and (-) with a wire link. When doing this, the

HP289

Image 53 Contents
USER’S Manual HP289 HFC Warranty Contents HP289 Introduction Model 214D- a Model Number DesignationIntroduction 2A Introduction 2B This page intentionally left blank HP289 General SpecificationPulse Output Operation Front Panel Operation Setting the Batch QuantityStarting a Batch ResettingStopping Displayed InformationAccumulated Total Limit on Batch SizeBatch Operations Page Control Relays Flow Alarm Signal TimeoutEnd-of-Batch End-of-Batch SignalAuto Restart Automatic Overrun Compensation Calculation of Rate and Total Frequency Input = fxHFiltering 103 For Example Total ConversionOutput Pulse and Flow Alarm Page Hardware OptionsRS232/422/485 Interface Option RS422 Interface HP289 Multipoint CommunicationRS485 Interface HP289 Printer Communication ProtocolComputer RUN ConfigurationPage Base Configuring the Setup ParametersStep Display Description Text Dis Signal Timeout in seconds Setting to Disables this featureEntering the Batch Parameters EOB Date Configuring the OptionsEcho Checking the Input Signal Input Circuit for the 214D Series Input CircuitsGeneral Specification Switch SettingsInput Circuits Frequency Input Circuit HP289 Page Page Remote Key Switches General InstallationRC Networks for Interference Suppression Terminal Relay Option & Switches Terminal Wiring DesignationsTerminalDescription Terminal RS232/422/485 OptionTrouble Shooting Instrument acting erratically No end of batch, pulse output, or flow alarmCounting erratically Input Errors Error CodesOutput Errors Configuration ErrorsFront Panel IndexTerminal Wiring