TRENDnet 18-1106-33 user manual Manual restart in operation mode Restart after power failure

Page 22

7. Detailed Description

7.5Manual restart in operation mode 8

Note: Although this feature activates the general cut-off, it is not a dedicated ”safety-function” of the type listed above. When the error message shown above is dis- played, the sensor must be exchanged by service per- sonnel.

Restart after power failure

MultiTemp III automatically restarts after a power failure. Make sure that this automatic restart will not cause a dange- rous situation to arise.

The unit can be restarted manually instead by pressing (see Section 7.5).

MultiTemp III is factory set to normal mode. Normal mode means that the unit automatically re-starts after power failu- re. As most modern electrophoresis power supplies automati- cally restart after short power failures, we strongly recom- mend that you leave MultiTemp III set in its normal mode.

However, if automatic re-start is inconvenient, the unit can be restarted manually by setting Operation mode 8 to ON, follow the procedure below:

Setting mode 8 to ON

1.Press simultaneously with the mains switch. Then

press again until ”mode 8” is selected.

2.Press to select ”ON” instead of ”OFF”. (”OFF” is factory set)

3.

Confirm by pressing

when the display is flas-

 

hing.

 

4.Wait until the display shows the actual temperature.

5.Switch off the unit using the mains switch.

The operation mode is now changed to ON.

20

Image 22
Contents MultiTemp Important user information Contents Page Safety precau- tions Safety informationIntroduction Safety standards Protection against over-heating Safety features General cut-offRestart after power failure Pump motor protectionInstallation Unpacking and installationInstallation Mains connection Operation Start-upSetting the temperature limit Tempera- ture setting limit Tempera- ture setting Tempera- ture correction value Operation Trouble shooting Error messagesGeneral MaintenanceDisposal Push Detailed Description Detailed Description Description of safety features General cut-off Accurate Setting of Temperature Limit Protecting the pump motorManual restart in operation mode Restart after power failure Setting mode 8 to onResetting mode 8 to OFF Technical specifications Technical specifications ambient temperature 20 CAccessories Quantity Code No Ordering informationPage Short instructions