Note: The following short instructions are intended as a guide to users who are fully familiar with safety procedures and operating instructions described earlier in this manual.

1.Check to make sure that all tubings are firmly connected and secured with hose clamps.

2.Check the liquid level and fill the bath if needed (the level should be 2-4 cm below the lid).

3.Switch the power on at the mains switch.

4.Set the safety temperature limit using the dial on the rear of the unit.

5.Set the setting range limit of the circulator (if required by the application or any safety precautions).

6. Set the working temperature. Press to select the S mode

.

Use or to set the working temperature. Confirm with

before the display automatically returns to the actual working temperature.

7.Set the temperature correction value if required.

A

Printed in Sweden by TK i Uppsala, July 1997

Image 27
Contents MultiTemp Important user information Contents Page Safety information Safety precau- tionsIntroduction Safety standards Pump motor protection Safety features General cut-offRestart after power failure Protection against over-heatingUnpacking and installation InstallationInstallation Mains connection Start-up OperationSetting the temperature limit Tempera- ture setting limit Tempera- ture setting Tempera- ture correction value Operation Error messages Trouble shootingMaintenance GeneralDisposal Push Detailed Description Detailed Description Description of safety features General cut-off Protecting the pump motor Accurate Setting of Temperature LimitSetting mode 8 to on Manual restart in operation mode Restart after power failureResetting mode 8 to OFF Technical specifications ambient temperature 20 C Technical specificationsOrdering information Accessories Quantity Code NoPage Short instructions