5. Trouble shooting

5. Trouble shooting

5.1Error messages

Message on display

Cause

Remedy

 

 

 

 

 

 

The bath temperature has exceeded

Pour in more liquid.

 

 

 

 

 

 

 

 

 

 

 

 

the safety temperature limit.

 

Too low water level.

The winding of the pump motor is too hot.

The temperature sensor is broken.

The bath liquid has cooled down below

the temperature limit.

The winding of the pump has cooled down below the alarm level.

The temperature sensor is

functioning again.

An undefined fault has occurred.

Pour in more liquid.

Check pump and wait a few minutes until the

motor temperature has dropped.

If this fault occurs often, return the unit for

service.

The sensor must be exchanged by service

personnel.

Check the water level in the bath and

adjust if necessary.Restart the unit by

pressing reset.

Check pump and wait a few minutes until the motor temperature has dropped.

If this fault occurs often, return the unit for

service. Restart the unit by pressing reset.

Restart the unit by pressing reset.

This error may require service personnel.

However, before contacting service personnel, top up the bath with some water. This can solve the problem.

Locked display

Undefined cause.

Press both arrow keys and the main switch

simultaneously to get in the display.

Switch off the unit and restart.

Note: The unit is now changed to the factory set parameters.

13

Image 15
Contents MultiTemp Important user information Contents Page Safety information Safety precau- tionsIntroduction Safety standards Pump motor protection Safety features General cut-offRestart after power failure Protection against over-heatingUnpacking and installation InstallationInstallation Mains connection Start-up OperationSetting the temperature limit Tempera- ture setting limit Tempera- ture setting Tempera- ture correction value Operation Error messages Trouble shootingMaintenance GeneralDisposal Push Detailed Description Detailed Description Description of safety features General cut-off Protecting the pump motor Accurate Setting of Temperature LimitSetting mode 8 to on Manual restart in operation mode Restart after power failureResetting mode 8 to OFF Technical specifications ambient temperature 20 C Technical specificationsOrdering information Accessories Quantity Code NoPage Short instructions