Extron electronic FPM 101, FLS 101 manual General

Page 27

Testing ranges

 

Measurement

-60 to +6 dBm

Tone detection

At 850 nm: -45 to +6 dBm

 

At all other wavelengths: -50 to +6 dBm

Wavelength identification

At 850 nm: -45 to +6 dBm

 

At all other wavelengths: -50 to +6 dBm

Resolution

0.01 dB

Accuracy

±0.25 dB at -10 dBm

General

Power

 

FLS 101

Internal batteries

 

or

 

External, supplied by optional AC adapter

 

Input: 100-240 VAC, 50-60 Hz

 

Output: 9 VDC, 1.5 A, 13 watts

FPM 101

Internal batteries

 

or

 

External

 

Supplied via USB port

Power consumption

 

FLS 101

LED transmission: 1.8 watts, 3 VDC

 

Laser transmission: 1.5 watts, 3 VDC

FPM 101

2.7 watts, 3 VDC with backlight on

Batteries

 

Number/type

2 AA batteries (1.5 V alkaline or 1.2 V rechargeable)

Operating time (nominal)

 

FLS 101

Multimode operation: 20 hours

 

Singlemode operation: 72 hours

FPM 101

300 hours

Temperature/humidity

Storage: -22 to +140 °F (-30 to +60 °C) / 10% to 90%, noncondensing

 

Operating: +14 to +122 °F (-10 to +50 °C) / 10% to 90%, noncondensing

Cooling

Convection, no vents

Thermal dissipation

1.1 BTU/hr

Mounting

No

Enclosure type

Plastic

Enclosure dimensions

4.5" H x 2.6" W x 1.0" D

 

(11.5 cm H x 6.7 cm W x 2.5 cm D)

 

(Excluding the rubber boot, connectors.)

 

5.5" H x 3.6" W x 1.5" D

 

(14.0 cm H x 9.1 cm W x 3.8 cm D)

 

(Including including rubber boot and connnectors.)

Product weight

 

FLS 101

0.7 lb (0.3 kg)

FPM 101

0.6 lb (0.3 kg)

Shipping weight

2 lbs (1 kg)

Vibration

ISTA 1A in carton (International Safe Transit Association)

Regulatory compliance

 

Safety

EN 61010-1

EMI/EMC

EN 61000-4, EN 61326-1

Warranty

3 years parts and labor, excluding batteries

NOTE: All nominal levels are at ±10%.

NOTE: Specifications are subject to change without notice.

Fiber Optic Test Meters • Reference 21

Image 27
Contents FPM 101 / FLS Safety Instructions English Conventions Used in this Guide Page Contents Fiber Optic Test Meters Contents Fiber Optic Test Set offers FeaturesAbout this User Guide About the FPM 101 and FLSPower Source FPM 101 Power Meter offersFLS 101 Light Source offers FLS 101 Light Source AccessoriesFPM 101 Power Meter Typical ApplicationsOperation FPM 101 Power Meter FPM 101 Power Meter Controls and ConnectionsPower Up the FPM Wavelength/Backlight ButtonDB/dBm/μW Button Splash and Auto-Off Screen Test Mode To Set a ReferenceTest Mode To View the Reference Ref/Set ButtonField Description Display Readings in Test ModeFLS 101 Light Source Controls and Connections FLS 101 PowerOutput Wavelength Selection Output Ports and IndicatorsTo Measure Optical Power Measuring Optical PowerTest the Transmit Reference Cable Testing Fiber Optic LinksFLS FPM Verify the Reference CablesFLS Insert the LinkConnections to Measure Link Insertion Loss To Quickly Identify a Fiber Optic Link Using the Tone GeneratorCleaning FPM 101 Optical Ports Cleaning Optical PortsBattery Replacement Cleaning FLS 101 Optical PortsRepair and Calibration Loops for Singlemode Transmit Reference Cables Mandrels for Multimode Transmit Reference CablesSpecifications Optical specifications FLS 101 fiber optic outputOptical specifications- FPM General Included Accessories Page USA, Canada, South America Japan Central America Extron Warranty