Calibrating the Scanner

 

Focus Calibration

The Flextight 848 scanner uses an adjustable zoom and focus mechanism to optimize its scanning resolution for each of the original formats it can handle. If you suspect your scans are not as sharp as they should be, then you may need to calibrate your scanner. Note that you must not calibrate the scanner every day. It is intended that the scanner is calibrated not more than once every three months. In most cases, this procedure will never be necessary.

Use procedure below to make Focus Calibration:

English

1.Locate the "Focus sheet" that came with your scanner. It is a 8 x 9 cm (3" x 3.5") square of transparent film with a black image on it.

Focus sheet

PN:40900026, rev. 1.10

2.Load the sheet into the 6 x 6 original holder with the lines pointing towards the scanner (vertically). Place the sheet as straight as possible. Use the scanners light table grid and the corners marked on the sheet to help align it.

3.Select Focus Calibration from the Maintenance menu.

4.Click on the Calibrate button, and the focus calibration starts.

5.When the focus calibration is finalized, the focus calibration window closes again.

Your scanner is now calibrated for all resolutions and for both transparent and reflective originals.

User's Guide - Flextight 848

Image 21
Contents Users Guide Users Guide Flextight Table of Contents English Hardware Reference Flextight English System Requirements Hardware Reference Flextight Users Guide Flextight Front Panel Front panel features of FlextightBack panel features of Flextight Back Panel & ConnectorsCut-away view of Flextight Cut-Away ViewSetting up the Scanner Environmental RequirementsElectrical Requirements Installation Procedure, FireWire Interface Operating the Flextight 848 scanner Operating Instructions English English Reflective holder, top Resolutions True ResolutionsOriginal Max. resolution ppi At full scanning width Calibrating the Scanner Use procedure below to make Focus Calibration Select Focus Calibration from the Maintenance menuFocus Calibration White Calibration for Reflectives Use procedure below to make White CalibrationAlign the top edge Replacing the Light Tubes Light Tubes Follow procedure below to replace a tubeGeneral Replacing the Light TubesEnglish Disposal Cleaning the External Surface of the ScannerTechnical Specifications FCC Notice Declaration of conformity CE Declaration of conformity