Fisher FB200-2 manual Automatic Power-fail Restart, Cleaning

Page 8

Enter a time parameter as described above for setting a voltage or current setpoint. Enter a time duration of 0-99 hours, 59 minutes. A time duration of zero effectively disables timed run mode.

When running a timed run, the time parameter displays the time remaining in the run. When running a non-timed run, the time parameter displays the elapsed time during the run

Automatic Power-fail Restart

Automatic power-fail restart capability allows a timed run which is interrupted by loss of AC power to be restarted automatically, so that the total time programmed for the time parameter will be met. NOTE: Loss of AC power includes switching off the unit using the power switch located at the rear of the unit. When performing timed runs with power-fail restart enabled, always use the front panel controls to stop a run in progress.

To enable power-fail restart, press the top function key when in the operating mode screen (constant voltage or current). The state of power-fail restart will toggle from off to on or on to off with each keypress.

When AC power is restored during a timed run in which power-fail restart is enabled, the display will show "POWER-FAIL RESTART" for approximately 10 seconds indicating power-fail restart pending. During this time the output is not energized, to allow safely stopping the run (by pressing Run). After the power-fail restart pending delay is completed, the power supply output will ramp up to the setpoint value and the time will continue from that point. Any number of power interruptions can occur during the completion of a single run.

Cleaning

Before cleaning the unit, be sure to always turn off AC power using the switch at the rear of the unit, and then disconnecting the AC line cord. The front of the unit is sealed and can be wiped clean with any mild detergent solution. Avoid harsh cleaners or agents as they may deteriorate the surface of the tactile membrane keys.

FB200-2 MANUAL

Page 7

7608175 Rev A 8/02

Image 8
Contents FB200-2 Electrophoresis Power Supply Operator’s ManualFisher Biotech FB200 Electrophoresis Power Supply Table of ContentsSafety Considerations Unpacking the Power Supply IntroductionSpecifications AC input PowerDC output Power EnvironmentalUsing the Power Supply Getting StartedTimed Operations Constant Voltage OperationsConstant Current Operations Automatic Power-fail Restart CleaningTroubleshooting and Error Indications Replacing a fuseFB200 Warranty Statement ComplianceTechnical Support Replacement PartsAccessories Table des matières Considérations de sécurité Déballage de l’alimentation électrique Spécifications Courant d’entrée CAEnvironnement Courant de sortie CCMise en route Utilisation de l’alimentation électriqueOpérations à tension constante Opérations à intensité constanteNettoyage Opérations à durée déterminéeRedémarrage automatique en cas de coupure de courant Remplacement d’un fusible Dépannage et indications d’erreurGarantie du FB200 ConformitéPièces de rechange Fisher Scientific Park Lane Pittsburgh, PA 15275 USAFisher Biotech FB200-2 Elektrophorese-Netzteil InhaltsverzeichnisSicherheitstechnische Hinweise Einführung Auspacken des NetzteilsTechnische Daten NetzeingangsleistungUmfeldbedingungen GleichstromausgangsleistungEinstieg Verwenden des NetzteilsBetrieb bei Dauerspannung Betrieb bei DauerstromReinigung Zeitgesteuerter BetriebAutomatischer Neustart bei Stromausfall Fehlersuche und Fehlermeldung Auswechseln einer SicherungFB200 Garantieerklärung KompatibilitätErsatzteile Contenido Consideraciones de seguridad Introducción Desembalaje de la fuente de alimentaciónEspecificaciones Potencia de entrada de CAMedioambientales Potencia de salida de CCPara empezar Uso de la fuente de alimentaciónOperaciones a voltaje constante Operaciones a corriente constanteLimpieza Operaciones temporizadasReinicio automático tras un fallo en la alimentación Diagnóstico e indicaciones de error Sustitución de un fusibleDeclaración de garantía del FB200 Conformidad Conformidad EMC Compatibilidad electromagnéticaPiezas de repuesto Fisher Scientific Park Lane Pittsburgh, PA 15275, EE.UUFB200-2 Electrophoresis Power Supply Part No Rev a