Fisher FB200-2 manual Troubleshooting and Error Indications, Replacing a fuse

Page 9

Troubleshooting and Error Indications

The FB200 detects and reports several events and conditions that are considered errors. The FB200 will stop any run in process and display "ERROR XX" where XX is one of the below listed errors. When the FB200 is displaying an error indication, press the gray ON/OFF key to clear the error and return to setup mode, or press the Run key to resume the run (after correcting the cause of the error). Note that loss of AC power during a run is considered an error condition. Always stop a run before turning off AC power to the FB200.

'XX'

Condition and possible remedy

00

Minimum load current detected. The output is not connected, or the

 

electrophoresis apparatus is not set up properly. Check your setup and

 

connections.

01

Maximum load current exceeded. The output is short circuit, or the

 

electrophoresis apparatus is not set up properly. Check your setup and

 

connections

02

Step load current change. The FB200 has detected a large change in load

 

current. Check your setup and connections

03

Loss of AC power; PF enabled and completed. The FB200 detected loss of

 

AC power, however the run was completed as programmed.

04

Loss of AC power; PF disabled and not completed. The FB200 detected loss

 

of AC power, the run was not completed since PF was not enabled.

05

Loss of AC power; run not timed. The FB200 detected loss of AC power.

06-99

The FB200 detected an internal error condition. Contact technical support for

 

instructions.

Before servicing the unit, be sure to always turn off AC power using the switch at the rear of the unit, and then disconnecting the AC line cord. The FB200 power supply requires no periodic servicing and should provide years of trouble free operation. Should you need to replace the fuses proceed as follows:

Replacing a fuse

CAUTION: The FB200 uses double pole neutral fusing.

Turn off AC power using the switch at the rear of the unit, and then disconnecting the AC line cord. Remove the fuse holder assembly using a small flat blade screwdriver. Always replace both fuses with appropriate replacement fuses: 3A, 250V, 5X20mm, type T fuse (T3A,250V) (Fisher catalog number FB-FUSE-2).

FB200-2 MANUAL

Page 8

7608175 Rev A 8/02

Image 9
Contents Operator’s Manual FB200-2 Electrophoresis Power SupplyTable of Contents Fisher Biotech FB200 Electrophoresis Power SupplySafety Considerations Introduction Unpacking the Power SupplyAC input Power SpecificationsDC output Power EnvironmentalGetting Started Using the Power SupplyConstant Voltage Operations Constant Current OperationsTimed Operations Cleaning Automatic Power-fail RestartReplacing a fuse Troubleshooting and Error IndicationsCompliance FB200 Warranty StatementReplacement Parts AccessoriesTechnical Support Table des matières Considérations de sécurité Déballage de l’alimentation électrique Courant d’entrée CA SpécificationsEnvironnement Courant de sortie CCUtilisation de l’alimentation électrique Mise en routeOpérations à intensité constante Opérations à tension constanteOpérations à durée déterminée Redémarrage automatique en cas de coupure de courantNettoyage Dépannage et indications d’erreur Remplacement d’un fusibleConformité Garantie du FB200Fisher Scientific Park Lane Pittsburgh, PA 15275 USA Pièces de rechangeInhaltsverzeichnis Fisher Biotech FB200-2 Elektrophorese-NetzteilSicherheitstechnische Hinweise Auspacken des Netzteils EinführungNetzeingangsleistung Technische DatenUmfeldbedingungen GleichstromausgangsleistungVerwenden des Netzteils EinstiegBetrieb bei Dauerstrom Betrieb bei DauerspannungZeitgesteuerter Betrieb Automatischer Neustart bei StromausfallReinigung Auswechseln einer Sicherung Fehlersuche und FehlermeldungKompatibilität FB200 GarantieerklärungErsatzteile Contenido Consideraciones de seguridad Desembalaje de la fuente de alimentación IntroducciónPotencia de entrada de CA EspecificacionesMedioambientales Potencia de salida de CCUso de la fuente de alimentación Para empezarOperaciones a corriente constante Operaciones a voltaje constanteOperaciones temporizadas Reinicio automático tras un fallo en la alimentaciónLimpieza Sustitución de un fusible Diagnóstico e indicaciones de errorConformidad Conformidad EMC Compatibilidad electromagnética Declaración de garantía del FB200Fisher Scientific Park Lane Pittsburgh, PA 15275, EE.UU Piezas de repuestoFB200-2 Electrophoresis Power Supply Part No Rev a