Agilent Technologies 8510-58 manual Modification procedure, Select standards

Page 7

Modification procedure

Calibration kit modification provides the capability to adapt to measurement calibrations in other con- nector types or to generate more precise error models from existing kits. Provided the appropri- ate standards are available, cal kit modification can be used to establish a reference plane in the same transmission media as the test devices and at a specified point, generally the point of device con- nection/insertion. After calibration, the resultant measurement system, including any adapters which would reduce system directivity, is fully cor- rected and the systematic measurement errors are mathematically removed. Additionally, the modifi- cation function allows the user to input more pre- cise physical definitions for the standards in a given cal kit. The process to modify or create a cal kit consists of the following steps:

1.Select standards

2.Define standards

3.Assign classes

4.Enter standards/classes

5.Verify performance

To further illustrate, an example waveguide cali- bration kit is developed as the general descriptions in MODIFY CAL KIT process are presented.

Select standards

Determine what standards are necessary for cali- bration and are available in the transmission media of the test devices.

Calibration standards are chosen based on the fol- lowing criteria:

A well defined response which is mechanically repeatable and stable over typical ambient tem- peratures and conditions. The most common coaxial standards are zero-electrical-length short, shielded open and matched load termina- tions which ideally have fixed magnitude and broadband phase response. Since waveguide open circuits are generally not modelable, the types of standards typically used for waveguide calibration are a pair of offset shorts and a fixed or sliding load.

A unique and distinct frequency response. To fully calibrate each test port (that is to provide the standards necessary for S11 or S22 1-PORT calibration), three standards are required that exhibit distinct phase and/or magnitude at each particular frequency within the calibration band. For example, in coax, a zero-length short and a perfect shielded open exhibit 180 degree phase separation while a matched load will pro- vide 40 to 50 dB magnitude separation from both the short and the open. In waveguide, a pair of offset shorts of correct length provide phase separation.

Broadband frequency coverage. In broadband applications, it is often difficult to find stan- dards that exhibit a known, suitable response over the entire band. A set of frequency-banded standards of the same type can be selected in order to characterize the full measurement band.

The TRL 2-PORT calibration requires only a sin- gle precision impedance standard—a transmis- sion line. An unknown high reflection device and a thru connection are sufficient to complete this technique.

7

Image 7
Contents Product Note 8510-5B Table of contents Introduction Measurement errorsMeasurement calibration Calibration kit Class assignment Standard definitionStandard definitions table Standard class assignments Select standards Modification procedureStandard definition models Define standardsStandard type Standard numberOpen circuit capacitance C0 , C1 , C2 and C3 ΠfZ ΠfXShort circuit inductance L0 , L1, L2 and L3 ∆∅radians = 2πf ∆lengthTerminal impedance Offset delayDelay seconds = Fixed or slidingOffset Z0 Linear delay Actual delay = Fco/f210 log10e εr Offset loss1GHz = 1GHz c Z0Lower/minimum frequency ∅radians = Upper/maximum frequencyFupper = 2 x Flower Coax or waveguideStandard labels Assign classesStandard Classes Isolation S11 A,B,C and S22 A,B,CForward transmission match and thru Reverse transmission match and thruTRM Thru TRL ThruTRL Reflect TRL LineTRL options Standard Class labelsCalibration kit label Verify performance Enter standards/classesModeling an arbitrary impedance standard User modified cal kits and Agilent 8510 specificationsModification examples Modeling a thru adapterTo store calibration kits from the Agilent 8510 onto a disk Appendix a Calibration kit entry procedureDisk procedure To load calibration kits from disk into AgilentFront panel procedure P-band waveguide example Pshort Mm coaxial connector interface Appendix B Dimensional considerations in coaxial connectorsType-N coaxial connector interface Female type-N Mm Coaxial connector Type-N coaxial connector interfacePage Equation Appendix C Cal coefficients modelTheir first order approximations, R is small and G=0, are Then Online assistance Phone or Fax United States