Agilent Technologies 8510-58 manual Enter standards/classes, Verify performance

Page 21

Again, cal kit labels should be chosen to best describe the calibration devices. The “B.1” default suffix corresponds to the kit’s mechanical revision

(B) and mathematical revision (1).

Note

To prevent confusion, if any standard definitions in a calibration kit are modified but a new kit label is not entered, the default label will appear with the last character replaced by a “*”. This is not the case if only a class is redefined without changing a standard definition.

The WR-62 waveguide calibration kit can be labeled simply – P BAND.

Enter standards/classes

The specifications for the Standard Definition table and Standard Class Assignments table can be entered into the 8510 through front panel menu- driven entry or under program control by an exter- nal controller. The procedure for entry of standard definitions, standard labels, class assignments, class labels, and calibration kit label is described in Appendix A.

Note

DO NOT exit the calibration kit modification process without saving the calibration kit defini- tions in the appropriate register in the 8510. Failure to save the redefined calibration kit will result in not saving the new definitions and the original definitions for that register will remain. Once this process is completed, it is recommended that the new calibration kit should be saved on tape.

Verify performance

Once a measurement calibration using a particular calibration kit has been generated, its performance should be checked before making device measure- ments. To check the accuracy that can be obtained using the new calibration kit, a device with a well defined frequency response (preferably unlike any of the standards used) should be measured. It is important to note that the verification device must not be one of the calibration standards. Calibrated measurement of one of the calibration standards is merely a measure of repeatability.

A performance check of waveguide calibration kits is often accomplished by measuring a zero length short or a short at the end of a straight section of waveguide. The measured response of this device on a polar display should be a dot at 1 ∠ 180°. The deviation from the known is an indication of the accuracy. To achieve a more complete verification of a particular measurement calibration, (including dynamic accuracy) accurately known verification standards with a diverse magnitude and phase response should be used. NBS traceable or Agilent standards are recommended to achieve verifiable measurement accuracy. Further, it is recommend- ed that verification standards with known but dif- ferent phase and magnitude response than any of the calibration standards be used to verify per- formance of the 8510.

21

Image 21
Contents Product Note 8510-5B Table of contents Measurement errors IntroductionMeasurement calibration Calibration kit Class assignment Standard definitionStandard definitions table Standard class assignments Select standards Modification procedureStandard definition models Define standardsStandard number Standard typeOpen circuit capacitance C0 , C1 , C2 and C3 ΠfZ ΠfXShort circuit inductance L0 , L1, L2 and L3 ∆∅radians = 2πf ∆lengthDelay seconds = Offset delayFixed or sliding Terminal impedanceOffset Z0 Linear delay Actual delay = Fco/f21GHz = Offset loss1GHz c Z0 10 log10e εrLower/minimum frequency Fupper = 2 x Flower Upper/maximum frequencyCoax or waveguide ∅radians =Assign classes Standard labelsStandard Classes Forward transmission match and thru S11 A,B,C and S22 A,B,C Reverse transmission match and thru IsolationTRL Reflect TRL ThruTRL Line TRM ThruStandard Class labels TRL optionsCalibration kit label Verify performance Enter standards/classesModification examples User modified cal kits and Agilent 8510 specificationsModeling a thru adapter Modeling an arbitrary impedance standardDisk procedure Appendix a Calibration kit entry procedureTo load calibration kits from disk into Agilent To store calibration kits from the Agilent 8510 onto a diskFront panel procedure P-band waveguide example Pshort Appendix B Dimensional considerations in coaxial connectors Mm coaxial connector interfaceType-N coaxial connector interface Female type-N Mm Coaxial connector Type-N coaxial connector interfacePage Equation Appendix C Cal coefficients modelTheir first order approximations, R is small and G=0, are Then Online assistance Phone or Fax United States