Agilent Technologies 8510-58 TRL Thru, TRL Reflect, TRL Line, TRM Thru, TRM Reflec, TRM Match

Page 19

TRL Thru

TRL Thru corresponds to the measurement of the S-parameters of a zero-length or short thru connec- tion between port 1 and port 2. The Thru, Reflect and Line classes are used exclusively for the three steps of the TRL 2-PORT calibration. Typically, a “delay/thru” with zero (or the smallest) Offset Delay is specified as the TRL Thru standard.

TRL Reflect

TRL Reflect corresponds to the S11 and S22 meas- urement of a highly reflective 1-port device. The Reflect (typically an open or short circuit) must be the same for port 1 and 2. The reflection coeffi- cient magnitude of the Reflect should be close to 1 but is not specified. The phase of the reflection coefficient need only be approximately specified (within ± 90 degrees).

TRL Line

TRL Line corresponds to the measurement of the S-parameters of a short transmission line. The impedance of this Line determines the reference impedance for the subsequent error-corrected measurements. The insertion phase of the Line need not be precisely defined but may not be the same as (nor a multiple of pi) the phase of the Thru.

TRM Thru

Refer to “TRL Thru” section.

TRM Reflec

Refer to “TRL Reflec” section.

TRM Match

TRM Match corresponds to the measurement of the S-parameters of a matched load. The input reflec- tion of this Match determines the reference imped- ance for the subsequent error-corrected measurements. The phase of the Match does not need to be precisely defined.

Adapter

This class is used to specify the adapters used for the adapter removal process. The standard num- ber of the adapter or adapters to be characterized is entered into the class assignment. Only an esti- mate of the adapter’s Offset Delay is required

(within ± 90 degrees). A simple way to estimate the Offset Delay of any adapter would be as fol- lows. Perform a 1-port calibration (Response or S11 1-PORT) and then connect the adapter to the test port. Terminate the adapter with a short cir- cuit and then measure the Group Delay. If the short circuit is not an offset short, the adapter’s Offset Delay is simply l/2 of the measured delay.

If the short circuit is offset, its delay must be sub- tracted from the measured delay.

Modifying a cal set with connector compensation Connector compensation is a feature that provides for compensation of the discontinuity found at the interface between the test port and a connector. The connector here, although mechanically com- patible, is not the same as the connector used for the calibration. There are several connector fami- lies that have the same characteristic impedance, but use a different geometry. Examples of such pairs include:

3.5mm / 2.92 mm

3.5mm / SMA SMA / 2.92 mm

2.4mm / 1.85 mm

The interface discontinuity is modeled as a lumped, shunt-susceptance at the test port refer- ence plane. The susceptance is generated from a capacitance model of the form:

C=C0 + C1  f + C2  f2 + C3  f3

where f is the frequency. The coefficients are pro- vided in the default Cal Kits for a number of typi- cally used connector-pair combinations. To add models for other connector types, or to change the coefficients for the pairs already defined in a Cal Kit, use the “Modifying a Calibration Kit” proce- dure in the “Calibrating for System Measurements” chapter of the 8510 network analyzer systems Operating and Programming Manual (part number 08510-90281). Note that the definitions in the default Cal Kits are additions to the Standard Class Adapter, and are Standards of type “OPEN.”

19

Image 19
Contents Product Note 8510-5B Table of contents Introduction Measurement errorsMeasurement calibration Calibration kit Class assignment Standard definitionStandard definitions table Standard class assignments Select standards Modification procedureStandard definition models Define standardsStandard type Standard numberOpen circuit capacitance C0 , C1 , C2 and C3 ΠfZ ΠfXShort circuit inductance L0 , L1, L2 and L3 ∆∅radians = 2πf ∆lengthTerminal impedance Offset delayDelay seconds = Fixed or slidingOffset Z0 Linear delay Actual delay = Fco/f210 log10e εr Offset loss1GHz = 1GHz c Z0Lower/minimum frequency ∅radians = Upper/maximum frequencyFupper = 2 x Flower Coax or waveguideStandard labels Assign classesStandard Classes Isolation S11 A,B,C and S22 A,B,CForward transmission match and thru Reverse transmission match and thruTRM Thru TRL ThruTRL Reflect TRL LineTRL options Standard Class labelsCalibration kit label Verify performance Enter standards/classesModeling an arbitrary impedance standard User modified cal kits and Agilent 8510 specificationsModification examples Modeling a thru adapterTo store calibration kits from the Agilent 8510 onto a disk Appendix a Calibration kit entry procedureDisk procedure To load calibration kits from disk into AgilentFront panel procedure P-band waveguide example Pshort Mm coaxial connector interface Appendix B Dimensional considerations in coaxial connectorsType-N coaxial connector interface Female type-N Mm Coaxial connector Type-N coaxial connector interfacePage Equation Appendix C Cal coefficients modelTheir first order approximations, R is small and G=0, are Then Online assistance Phone or Fax United States