The array coefficients are computed by measuring a set of “known” devices connected at a fixed point and solving as the vector difference between the modeled and measured response.

The full 2-port error model shown in Figure 1 is an example of only one of the measurement calibra- tions available with the 8510. The measurement calibration process for the 8510 must be one of seven types: RESPONSE, RESPONSE & ISOLATION, Sl l 1-PORT, S22 1-PORT, ONE PATH 2-PORT, FULL

2-PORT, and TRL 2-PORT. Each of these calibration types solves for a different set of the systematic measurement errors. A RESPONSE calibration solves for the systematic error term for reflec- tion or transmission tracking depending on the S-parameter which is activated on the 8510 at the time. RESPONSE & ISOLATION adds correction for crosstalk to a simple RESPONSE calibration. An S11 l-PORT calibration solves for the forward error terms, directivity, source match and reflec- tion tracking. Likewise, the S22 1-PORT calibration solves for the same error terms in the reverse direction. A ONE PATH 2-PORT calibration solves for all the forward error terms. FULL 2-PORT and TRL 2-PORT calibrations include both forward and reverse error terms.

The type of measurement calibration selected by the user depends on the device to be measured (i.e., 1-port or 2-port device) and the extent of accuracy enhancement desired. Further, a combi- nation of calibrations can be used in the measure- ment of a particular device.

The accuracy of subsequent test device measure- ments is dependent on the accuracy of the test equipment, how well the “known” devices are mod- eled and the exactness of the error correction model.

Calibration kit

A calibration kit is a set of physical devices called standards. Each standard has a precisely known or predictable magnitude and phase response as a function of frequency. In order for the 8510 to use the standards of a calibration kit, the response of each standard must be mathematically defined and then organized into standard classes which corre- spond to the error models used by the 8510. Agilent currently supplies calibration kits with

1.0-mm (85059A), 1.85-mm (85058D), 2.4-mm (85056A/D/K), 3.5-mm (85052A/B/C/D/E), 7-mm (85050B/C/D) and Type-N (85054B) coaxial con- nectors. To be able to use a particular calibration kit, the known characteristics from each standard in the kit must be entered into the 8510 non- volatile memory. The operating and service manu- als for each of the Agilent calibration kits contain the physical characteristics for each standard in the kit and mathematical definitions in the format required by the 8510.

Waveguide calibration using the 8510 is possible. Calibration in microstrip and other non-coaxial media is described in Agilent Product Note 8510-8A.

4

Image 4
Contents Product Note 8510-5B Table of contents Introduction Measurement errorsMeasurement calibration Calibration kit Standard definition Class assignmentStandard definitions table Standard class assignments Modification procedure Select standardsDefine standards Standard definition modelsStandard type Standard numberOpen circuit capacitance C0 , C1 , C2 and C3 ΠfX ΠfZ∆∅radians = 2πf ∆length Short circuit inductance L0 , L1, L2 and L3Offset delay Delay seconds =Fixed or sliding Terminal impedanceLinear delay Actual delay = Fco/f2 Offset Z0Offset loss 1GHz =1GHz c Z0 10 log10e εrLower/minimum frequency Upper/maximum frequency Fupper = 2 x FlowerCoax or waveguide ∅radians =Standard labels Assign classesStandard Classes S11 A,B,C and S22 A,B,C Forward transmission match and thruReverse transmission match and thru IsolationTRL Thru TRL ReflectTRL Line TRM ThruTRL options Standard Class labelsCalibration kit label Enter standards/classes Verify performanceUser modified cal kits and Agilent 8510 specifications Modification examplesModeling a thru adapter Modeling an arbitrary impedance standardAppendix a Calibration kit entry procedure Disk procedureTo load calibration kits from disk into Agilent To store calibration kits from the Agilent 8510 onto a diskFront panel procedure P-band waveguide example Pshort Mm coaxial connector interface Appendix B Dimensional considerations in coaxial connectorsType-N coaxial connector interface Mm Coaxial connector Type-N coaxial connector interface Female type-NPage Appendix C Cal coefficients model EquationTheir first order approximations, R is small and G=0, are Then Online assistance Phone or Fax United States