6.2.2Sensor/Electronics Functionality Verification

1.Turn power off to Innova-Switch.

2.Allow a 5 minute cool down.

3.Measure the resistance of each RTD at pins 1 and 6 of TBA (see Figure 6.0 or 7.0) for the hot RTD and pins 3 and 5 of TBA for the cold RTD. These resistances should be 110 ± 10 ohms (with sensors at approximately 70°F) and within 5% of each other in value.

4.Measure the insulation resistance between pin 1 of TBA and the case of the Innova-Switch. It should greater than 20 megohms.

5.If the Innova-Switchsensor assembly resistances are not as specified above, the switch sensor assembly must be replaced.

6.If the Innova-Switchsensor assembly resistances are as specified, the Innova- SwitchPC electronic board must be replaced.

SPECIAL NOTICE

The electronic assemblies contained in the Innova-Switch™ are configured for specific voltages and have specific modifications to accommodate the various agency approvals. When ordering spare electronics, replacements, or exchanges in the field please ensure you identify the specific configuration you have by noting the boxes marked on the transformer configuration tag.

IM-215 Rev-A.1

Series Innova-Switch™

Page 31 of 42

Image 31 Contents
INNOVA-SWITCH Series Before Starting Special Notice This manual covers the following model numbersTable of Contents Introduction Description Description IntroductionIM-215 Rev-A.1 Series Innova-Switch IM-215 Rev-A.1 Series Innova-Switch Thermal Differential Theory of Operation Level SwitchingFlow Switching Gas Or Liquid Flow Innova-Switch Model FS4200 Flow Response for Three Media IM-215 Rev-A.1 Series Innova-Switch Innova-Switch Models with the CE Option Switch Block Diagram Innova-Switch Switch Assembly When Installing Your Sierra Instruments Switch Into InstallationPipe or Vessel USE a 1 1/8 Inch 28.575mm OPEN-END IM-215 Rev-A.1 Series Innova-Switch Innova-SwitchSWITCH Local Electronics Field Wiring Diagram Electrical InstallationIM-215 Rev-A.1 Series Innova-Switch Wiring Innova Switch Flow Switch Remote Electronics Option FieldRemote Electronics Cable Termination ConnectionsInnova-SwitchSWITCH Remote Electronics Field Wiring Diagram L.E.D. and Relay Status Logic Fail-safe Innova-Switch Electronics For Optimum OPERATION, Calibration Must be Accomplished Pre-Operational Check RED Green Relay Contact Sensor Status LED Innova-Switch Electronics Calibration Level IM-215 Rev-A.1 Series Innova-Switch IM-215 Rev-A.1 Series Innova-Switch Cleaning Intenance and Trouble ShootingTroubleshooting Special Notice Process Connections Specifications TypeOperating Temperature Input PowerWarranty Warranty and ServiceService Connector Plates To Convert from Multiply by Volume Flow Conversion ChartFlow Conversion Chart Flow of Water Through Schedule 40 Steel Pipe Options Livetap LT Variable Insertion Thermocouple Output to and 10.4 RTD Output RT IM-215 Rev-A.1 Series Innova-Switch