3.0Installation
3.1Mechanical Installation
The standard
*IMPORTANT*
WHEN INSTALLING YOUR SIERRA INSTRUMENTS SWITCH INTO
A PIPE OR VESSEL USE A 1 1/8 INCH (28.575mm) OPEN-END
WRENCH TO TIGHTEN AT THE HEX FLATS OF THE MNPT OF A
STANDARD SWITCH. (IF YOU HAVE A
AN ALTERNATE SIZE WRENCH MAY BE REQUIRED). DO NOT
USE THE INSTRUMENT HEAD TO TIGHTEN THE SWITCH TO THE
MOUNTING PORT. ROTATION OF THE INSTRUMENT HEAD WITH
RESPECT TO THE SENSOR BODY CAN CAUSE INTERNAL
WIRING DAMAGE.
*IMPORTANT*
THE SWITCH BODY MUST BE ORIENTED TO HAVE THE TWIN SENSORS PROPERLY ORIENTED. DUE TO THE PIPE THREAD MOUNTING, IT MAY BE NECESSARY TO MAKE A TRIAL FIT, ADD OR REMOVE TEFLON TAPE OR OTHER PIPE THREAD SEALANT, AND REINSTALL TO ACHIEVE A SATISFACTORY SEAL WITH THE SENSORS PROPERLY ORIENTED. PROPER ORIENTATION IS MARKED ON THE SWITCH BODY FOR REFERENCE. SEE FIGURE 5.0 FOR DETAILS.
| Series | Page 15 of 42 |