Omega DPS3200, DPS3100 manual Channel Configuration, Setup for Voltage or Current Input

Page 11

CHANNEL CONFIGURATION

After entering correct pass-code and selecting the blinking 'CH' (ref. SETUP section), the display shows 'SLCt CH' (for "Select Channel"). Use ^v key to display the desired channel. Selected channel will be displayed in the format 'CHANEL x' (where x=channel#). Once the desired channel # is displayed, push SETUP key to go on to setup parameters for the displayed channel #.

Setup For Input Type

The display will briefly show 'IP TYPE' (for "Input type") after which the present input type for the selected channel will be displayed. Various inputs available are: J T/C, CR.AL T/C (Chromel Alumel, type K T/C), T T/C, E T/C, CURRENT, or VOLTAGE. If the unit has a noble metal thermocouple option, then only one T/C can be selected i.e. B T/C, R T/C, or S T/C. Use the ^v key to display the desired input type. Once the desired input type is displayed, push the SETUP key to enter that input type and go to Setup For Channel ON/OFF.

NOTE : If you have a millivolt input unit (instead of thermocouple), the various

types of inputs selectable are CURRENT, HI VOLT (0-10Vdc input) and LO VOLT (for millivolt input).

Setup for Voltage or Current Input:

Setup For Channel ON/OFF

After selecting input type, the next parameter for configuration is Channel ON/OFF, which determines whether a channel is scanned or not. If for any reason a channel is not being used, it must be turned OFF. In this way the unit will not spend any time scanning it. This will also prevent the unit from displaying a 'HELP' message on a channel that has nothing hooked up to it. Use ^v key to select the desired setting. If a channel is OFF, the display will show 'CHx OFF' , and if the channel is ON the display will show 'CHx ON' (x=channel #).

Use ^v key to display the desired ON/OFF setting, and then push SETUP key to enter that setting and go on to setup Decimal Point Position.

Setup For Decimal Point Position

After turning a channel ON or OFF, the next parameter for configuration is setting the decimal point position. This is indicated by the display showing ' 9999' . ^v key moves the decimal point through all the possible positions. After '9.999', the display goes to '9999' which indicates a display with no decimal point. Use ^v key to move the decimal point to the desired position, and then push SETUP key to enter that setting and go to setup High Scale.

Setup For High Scale

After setting a channel's decimal point position, the next parameter to be set is Hi Scale. This parameter determines what number will be displayed when the transducer puts out its maximum signal. For example, suppose a pressure transducer produces a 0Vdc to 10Vdc signal which corresponds to 0psi to 150psi. Then you would want the display to show 150 when 10Vdc is measured. To do this, set the Hi Scale=150.

After pushing SETUP key, the display will briefly read 'HI SCLE' (for "High Scale"), and then show the current High Scale setting. The flashing digit is the active digit. Pushing ^v key and keeping it pushed, will increment the digit. Releasing ^v

Page 10

Image 11
Contents DPS3100 & DPS3200 series Programmable Scanners Display Modes SetupChannel Configuration System ConfigurationError Messages and Solutions Mounting Table of ContentsTrouble Shooting Manual File omga31320404 SpecificationsRTD DescriptionDPS3100 & DPS DPS3200 To Display Perform Following Steps How to Display ParametersDisplaying Parameters Display Modes Elapsed Time ModeScan Mode Mode Function PerformedTypes of Deviation Displays High/Low Channel ModeDeviation Mode DPS3000 Series is SetupDifference Mode Display Time Calibration/Range settingCH Parameter List Display Options Relay Latch/Non-latchSetup for Voltage or Current Input Channel ConfigurationSetup For Setpoint Setup For Low ScaleSetup For Offset Setup For TareSetup For Relay Normally Open/Closed Setup for Universal Relay Channel 1 RelaySetup For Limits Setup For DeadbandSetup Examples Setup for Thermocouple/ Thermistor/ RTD InputsSetup For Display Time System ConfigurationSetup For Display Options Setup For Channel Calibration Setup For Relay Latch/Non-LatchSetup for Buzzer Setup For Cold Junction CalibrationCorrect Range Settings Thermocouple Calibration ProcedureVoltage Range Setup for millivolt inputs Current Range SetupPeaks Programming Flow ChartsChannel Diff Enter PasscodeSYS CH Select ChannelRelay Deadband Relay NO/NC Engineering Unit Thermistor LO VoltageCH ‘X’ Setpoint CH1 RELAY/UNIVERSAL Relay CH ‘X’ LimitOpen Collector Option RateLimits OutputsFIG- 3. Connections & Rear View of Instrument Power Pin # Signal Relay Output Open Collector Output Connector PIN AssignmentFactory Channel settings Trouble ShootingError Messages and Solutions Mounting Panel Cutout and mounting hole dimensionsWarranty Page