Omega DPS3200, DPS3100 Saving Parameter, Thermocouple Calibration Procedure, Current Range Setup

Page 17

desired value is displayed, push SETUP key to enter that value and go to the next step.

Voltage Range Setup (for millivolt inputs)

For Millivolt input units, the display first shows 'HV rnGE' (for "High Voltage Range"). After the High Voltage range is set, the display shows 'LV rnGE' (for "Low Voltage Range"). Enter values for respective inputs.

Current Range Setup

After VOLTAGE range has been set, the CURRENT range may be set. The display will briefly show 'C RANGE' (for "Current Range") followed by the current setting. Use ^v key (ref. Setup For High Scale) to set the desired CURRENT value. Once the correct CURRENT range is displayed, push SETUP key to enter that value and return to 'SYS CH' display.

Correct Range Settings

The following RANGE values should be entered for various inputs:

INPUT

RANGE VALUE

0-5 Vdc

5.000

0-10

Vdc

10.000

0-100 Mv

100.0

4-20

Ma. (loop current)

20.00

Thermocouple Calibration Procedure

Note: Make sure the unit is reading correct cold-junction temperature before calibrating.

If incorrect, adjust as described in the "Setup For Cold Junction" section.

For a thermocouple channel calibration (type J T.C), following steps should be performed. Note that calibrating any channel automatically sets the calibration for all channels. Also, only one type of thermocouple input needs to be calibrated i.e. J,K,T or E. For example, if the calibration is done for a type K thermocouple, types J, T, and E are automatically calibrated.

1.Connect a thermocouple calibration source to channel 1.

2.Dial in 1100 degrees centigrade (Note: unit must be programmed for displaying in centigrade).

3.Adjust the gain pot on the back of the instrument until the display reads '1100'.

4.Short channel #1's input with a wire or shorting bar.

5.Push DIFF key. The display will read '0002' --- or some other value.

6.Adjust offset pot on back of instrument (ref. Fig. 3) until the display reads '0000'.

7.Push SETUP key -- the display will show Cold Junction temperature.

8.Remove the shorting bar from channel one input and connect the thermocouple calibrator again.

9.Repeat steps 2 through 7 till the unit reads proper temperature.

SAVING PARAMETER:

DPS3000 series saves all the programmed parameters in an EEPROM (electrically erasable programmable read only memory). An EEPROM stores the parameters even when power is removed from the unit. However, it is important to note that if the parameters are changed during setup, they must be saved in the EEPROM by pressing reset key as described under setup. If the parameters are not saved and the power is removed from the unit, any newly changed values will be lost ( the unit will, however, maintain the old values).

Page 16

Image 17
Contents DPS3100 & DPS3200 series Programmable Scanners Channel Configuration SetupSystem Configuration Display ModesError Messages and Solutions Mounting Table of ContentsTrouble Shooting Manual File omga31320404 SpecificationsRTD DescriptionDPS3100 & DPS DPS3200 To Display Perform Following Steps How to Display ParametersDisplaying Parameters Scan Mode Elapsed Time ModeMode Function Performed Display ModesTypes of Deviation Displays High/Low Channel ModeDeviation Mode DPS3000 Series is SetupDifference Mode CH Parameter List Calibration/Range settingDisplay Options Relay Latch/Non-latch Display TimeSetup for Voltage or Current Input Channel ConfigurationSetup For Offset Setup For Low ScaleSetup For Tare Setup For SetpointSetup For Limits Setup for Universal Relay Channel 1 RelaySetup For Deadband Setup For Relay Normally Open/ClosedSetup Examples Setup for Thermocouple/ Thermistor/ RTD InputsSetup For Display Time System ConfigurationSetup For Display Options Setup for Buzzer Setup For Relay Latch/Non-LatchSetup For Cold Junction Calibration Setup For Channel CalibrationVoltage Range Setup for millivolt inputs Thermocouple Calibration ProcedureCurrent Range Setup Correct Range SettingsPeaks Programming Flow ChartsSYS CH Enter PasscodeSelect Channel Channel DiffCH ‘X’ Setpoint CH1 RELAY/UNIVERSAL Relay Thermistor LO VoltageCH ‘X’ Limit Relay Deadband Relay NO/NC Engineering UnitLimits RateOutputs Open Collector OptionFIG- 3. Connections & Rear View of Instrument Power Pin # Signal Relay Output Open Collector Output Connector PIN AssignmentFactory Channel settings Trouble ShootingError Messages and Solutions Mounting Panel Cutout and mounting hole dimensionsWarranty Page