Chapter 6 Error Messages / Analog Output Calibration
6.1 Error Messages
The DPF500 is provided with extensive self checking which assists the user in the location of setup entry errors and in reporting malfunctions or unusual operating conditions. When an error occurs, the display will flash. Press any key to see the error message corresponding to the error that has occurred. Press any key again to acknowl- edge the error. (If the error can be eliminated by a change of setup values, the unit will automatically advance to the MENU so that the appropriate setup changes can be made).
Table - 1 illustrates the warning message, problem, and recommended corrective actions.
Diagnostic Error Messages |
|
|
WARNING MESSAGE | CAUSE | CORRECTIVE ACTION |
RATE ER | Rate Low set | Set Rate Hi greater |
| higher than Rate Hi | than Rate Lo |
FAC ERR | Factor = 0 | Set in correct Factor |
“BAT” Descriptor | Low Battery | Replace battery(ies) |
E total | Total rollover | None required |
E RATE | Rate exceeds 99999 | Use lower rate dec point |
E pulse | Pulse out Overfl ow | Use different pulse scaler or |
|
| totalizer decimal point |
E FLASH | Save to fl ash | Write down displayed total and |
| memory failed | setup values if you are changing |
|
| the battery. If total wasn't saved, |
|
| it will display an arbitrary total |
|
| when new battery is installed. In |
|
| this case, reset the total to 0 and |
|
| check the setup information. |
6.2 Analog Output Calibration
CAUTION: Performing the analog output calibration will erase all programming values. Please record all programmed values before beginning the analog output calibration
If the unit is equipped with the analog output option, the
The
If the output is out of calibration, perform the following:
Remove power and/or disconnect battery(ies). Hold down the scroll ( ↑ ) key and replace the battery. This will initiallize the unit and advance to the analog output calibration mode. Connect a DMM set to read current in series with a power supply (8.5 to 30 volts DC) to TB12 (+) and TB11
If it does not, Adjust the numbers on the display up or down until the output reads 4.000 ma (± 0.005). Press the "E" key. The DMM should now read 20.000 ma (± 0.005). If it does not, Adjust the numbers on the display up or down until the output reads 20.000 ma (± 0.005). Press "E" and the unit will return to the "RUN" mode.
11