Omega DPF500 Series manual Typical Applications / Wiring Diagrams

Page 8

2.4 Wiring

Several typical applications of the DPF500 are shown below. Please observe that the various pulse inputs and power options may be intermixed in many ways to solve common applications. The isolated pulse output may be freely used so long as proper polarity is observed.

Caution : When 4-20 mA loop option is provided, the power wiring to the loop power option should always be to terminals (+) 12 and (-) 11. Accidental wiring to (+) 12 and (-)6 should be avoided since excessive current fl ow may result.

Caution : The magnetic pickup input and contact closure input require isolated sensors for proper operation. Ac- cidental connections to earth may result in erroneous operation of the analog output and/or excessive current fl ow.

Caution : Accidental connections from circuit common (3 or 6) to earth or terminal (11) may result in erroneous op- eration of the analog output and/or excessive current fl ow.

2.5 Typical Applications / Wiring Diagrams

CONTACT INPUT / PULSE OUTPUT / BATTERY POWERED

ISOLATED INPUT / 4-20mA LOOP POWERED

 

121110

32 1

 

9 8 7

6 5 4

4

1

 

123456

 

DPF76 Totalizer

Flowmeter with

 

 

Switch Closure Output

To Shield

Pin3

- +

24 VDC

Power

Supply

FlowmeterPSU-24B with DC Pulse Output

+ -

121110 9 8 7

6 5 4 3 2 1

+-

Strip Chart

Recorder

+24PowerVDC - Supply

PSU-24B

4-20mA (+)/DC In (+) 12

1

Mag Input1

4-20mA (-) 11

2

Mag Input2

Opto Input (+) 10

3

Shield/GND

Opto Input (-) 9

4

Reset Input

Opto Out (+) 8

5

Contact Input

Opto Out (-) 7

6

Common/ DC In (-)

MAG INPUT / DPF5-BP POWERED

DPF5-BP

DPF5-BP PLUG-IN

 

CONNECTOR

121110

3 2 1

9 8 7

6 5 4

 

Remote

Reset Switch

4-20mA (+)/DC In (+) 12

1

Mag Input1

 

4-20mA (-) 11

2

Mag Input2

 

Opto Input (+) 10

3

Shield/GND

 

Opto Input (-) 9

4

Reset Input

Turbine Meter with

Opto Out (+) 8

5

Contact Input

Mag Pickup

Opto Out (-) 7

6

Common/ DC In (-)

 

 

 

 

4-20mA (+)/DC In (+) 12

1 Mag Input1

 

 

 

 

 

4-20mA (-) 11

2

Mag Input2

NOTE: Separate power supplies

 

 

Opto Input (+) 10

3 Shield/GND

are recommended to isolate output

 

Opto Input (-) 9

4

Reset Input

from input.

 

 

 

 

Opto Out (+) 8

5 Contact Input

If isolation is not a concern, use a

 

Opto Out (-) 7

6 Common/ DC In (-)

single power supply

 

 

 

 

 

 

 

MAG INPUT / DC POWERED

 

15 VDC +

121110

 

 

32 1

 

 

Power

 

 

 

 

 

Supply

-

9

 

 

4

 

 

FPW-15

8

 

 

5

 

 

 

7

 

 

6

 

 

 

 

 

 

 

Remote

 

 

 

 

 

 

 

Reset Switch

 

4-20mA (+)/DC In (+) 12

1 Mag Input1

 

 

 

 

4-20mA (-) 11

2

Mag Input2

 

 

 

Opto Input (+) 10

3 Shield/GND

 

 

 

 

Opto Input (-)

9

4

Reset Input

 

Turbine Meter with

 

 

Opto Out (+)

8

5 Contact Input

 

Mag Pickup

 

 

Opto Out (-)

7

6 Common/ DC In (-)

 

4

Image 8
Contents User’sGuide Servicing North America Table of Contents Introduction Chapter InstallationChapter Error Messages/Analog Output Calibration Chapter Point Linearization OperationThis page Intentionally Blank Features Available Options/AccessoriesIntroduction Unpacking DescriptionInstallation Battery Installation Battery ReplacementDimensions DPF5-BP Accessory DPF502DPF501 DPF503Wiring Typical Applications / Wiring DiagramsProgramming Explanation of Display Prompts FT3Explanation of Display Prompts Record this number for later use!LOCK Code Programming Flowchart Rate234 Maintenance Operation GeneralError Messages / Analog Output Calibration Analog Output CalibrationAccuracy WARRANTY/DISCLAIMER Shop online at omega.com