Hach DR 5000 user manual Connection, Installation

Page 14

Installation

3.3 Connection

The DR 5000 has three USB interfaces as a standard feature, located on the back of the instrument (Figure 1).

The USB Type A interfaces are used for communications with a printer, USB memory stick or keyboard. A USB memory stick is used to update instrument software.

The USB Type B interface is used for communications with a PC. The optional Hach Data Trans software (see section 5.2.6.3 on page 31) must be installed on the PC for this use.

A USB hub may be used to connect several accessories at a time.

Note: USB cables must not be longer than 3 meters (10 feet).

These USB interfaces enable data and graphics to be output to a Printer and a PC and upgrade of instrument software (see section 6.8.2 on page 106).

Figure 1 Interfaces

1

USB type A

4

Fuse

 

 

 

 

2

USB type B

5

Plug in power supply

 

 

 

 

3

On/Off switch

 

 

 

 

 

 

14

Image 14
Contents January 2008, Edition Page Table of Contents Table of Contents 101 Table of Contents Section Specifications Specifications General Information Safety informationPrecautionary labels Use of hazard informationGeneral Information Class 1 Laser Chemical and Biological SafetySource Lamp Safety Overview of product General InformationGeneral Information Operating environment Section InstallationUnpack the instrument Installation ConnectionCell compartment 1 100 mm adapter optional Multi Cell Holder Top and BottomInstallation Installation Multi Cell Holder Installation Multi Cell HolderBeam path Beam pathStart Up Power the instrument on and offChanging the language setting Language selectionStart Up Tips for the use of the touch screen Standard OperationsOverview Use of the alphanumeric keypadStandard Operations Main Menu Main Menu optionsOperator ID Instrument Setup modeStandard Operations Standard Operations Sample ID Date and time Display and sound preferencesLamp Control VIS-Lamp and UV-LampEffect caused by the lamp settings on the Measuring Mode Standard Operations Standard Operations PC and printer Printer setupUSB connector KeyboardPrinter Setup Print data Hach Data TransPassword Password deactivation Standard Operations Auto/manual data storage Store, recall, send and delete dataData Log Recall stored data from the data logSend data from the data log To send measurement data to a PCDelete stored data from the data log Select Wavelength Scan or Time Course to recall data Data storage from wavelength scan or time courseRecall stored data from wavelength scan or time course Send data from wavelength scan or time course Option Press Wavelength Scan or Time Course and then Delete stored data from wavelength scan or time course Stored program options Stored ProgramsSelect a saved test/method enter user-specific basic data Stored programs optionsStandard Adjust Timer iconReading Mode Options Description Store Off/OnStandard Operations Use of program timers Set the reading modeTake single wavelength measurements continuous readings Set the dilution factorTake single wavelength measurements single reading Press OptionsMore…Dilution FactorRun a standard adjust Set the chemical form Change of the default setting of the chemical formPress Save as Default Press OptionsMoreChemical FormPress OptionsMoreReagent Blank Standard Operations Run a reagent blankMeasurement/analysis of a reagent blank Standard Operations Analysis of samples Add stored programs to the favorite programs listBarcode Programs Cell compartment #1Select the measuring range Changing the standard settingStandard Operations Complete a barcode 13 mm test/vial Select the chemical evaluation formBasic test-specific and sample-specific data settings Barcode program optionsEdit Sample blankOptions Description Send Data icon Program update options Update/edit barcode testsManual update of a barcode test Enter the NameEnter the Version Enter the Measurement ProcessTimer 1, Timer 2, Timer 3 and Timer Option 2 Manual check/revision of test data Update an existing barcode testOption 1 Manual update of test data Press OptionsMoreEditUpgrade of the instrument software Program a new testPress OK to return to the System Checks menu Standard Operations Program a user method Advanced OperationsUser Programs Program Name Advanced OperationsProgram Number Program TypeUnits Single wavelength settingsConcentration resolution number of decimal places Wavelength single wavelength program typeAbsorbance formula multi-wavelength program type Multi wavelength settingsChemical form Concentration Factor Kx List of available absorbance formulasWavelength λx Calibration by entering calibration values Calibration settings for single and multi wavelength modeCalibration settings Advanced Operations Advanced Operations Advanced Operations Upper and lower limits of the measuring range Store a user programAdditional user-defined parameters and functions Timer 1 / Timer 2 / Timer 3 / Timer Chemical form 2 / chemical form 3 / chemical formDefinitions of the program points Advanced Operations Free programming program typeMeasurement process Content and definition of the keys Key / Zeroing Enter a new element of a measuring sequenceProcess Timer key Key / ReadingInsert an element of a measuring sequence Enter the calibration formula evaluation formulaDelete an element of a measuring sequence Enter calibration formula C1New Number Edit formula key descriptionsScreen Key Description Enter the next calibration formula C2 or Cn C2 OnSelect Conc. key Enter variablesSave a free programming user program Advanced Operations Select a user program 4 Add, edit and delete user programs from the favorites listDelete Add to FavoritesEdit Delete a favorite program Favorite ProgramsRecall a favorite program Method Measures to identify other interference factors ChecklistStandard Addition monitoring/checking results Check the standard solution Peak volume Standard additionSample volume Complete a standard additionPeak volume/sample volume methods Press OptionsMoreStandard AdditionDescription of the table of readings Advanced Operations Program Set up single wavelength modeSingle wavelength setup options Options Description Concentration resolutionConcentration factor Take single wavelength measurements continuous readings Reading modeTake single wavelength measurements single reading Set the reading mode at different wavelengths Multi-wavelength setup optionsAbsorbance formula Press Absorbance Formula Concentration factor For data storage, see .3.1.1 on Set up the wavelength scan Wavelength scan setup optionsScale & Units Setting wavelengthOption Description Integral Cursor modeSelect view displaying table Wavelength scan reading Scale & units Press Scale & UnitsNavigating the wavelength scan Advanced Operations Work with reference scans Highlight Reference100 Time course of absorbance/transmittance Time course setup parametersTime course setup options 101Scale & units Time course scan readingTime & interval 102Analysis of time course data 103104 Navigation of a time scan or a time scan analysisNavigating the time scan 105 System checksInstrument information 106 Advanced Operations Upgrade of the instrument softwareOptical checks Wavelength check 107Noise check 108109 Stray light check 110Absorbance check 111Drift check 112Verification kit Press Nominal Values 113Press Verification 114115 Output checksLamp history 116 Advanced Operations Factory serviceService time Instrument Backup 117Restore backup data 118119 120 Spectrophotometer MaintenanceCleaning requirements DisplayTungsten lamp replacement Lamp replacementMaintenance 122Deuterium lamp UV replacement 123Fuse replacement 124125 Filter pad maintenanceFilter pad replacement Base of DR 5000 with filter grid 126Section Troubleshooting 127Troubleshooting 128Description Cat.-No Section Replacement Parts129 Replacement Parts 130Contact Information 131Contact Information 132Limitation of Remedies Limited WarrantyLimitations 133Limited Warranty 134Index 135136