Hach DR 5000 user manual Set up the wavelength scan, Wavelength scan setup options

Page 94

Advanced Operations

6.6Wavelength Scan mode – recording of absorbance and transmittance spectrums

In the wavelength scan mode, the absorbance of the light in a solution over a defined wavelength spectrum is measured.

The measurement results can be displayed as a curve, as percent transmittance (%T) or as absorbance (Abs). The collected data can be printed as a table or a curve.

The data are available for formatting changes. These include automatic scaling and zoom functions. Maximum and minimum values are determined and shown as a table.

The cursor can be moved to any point on the curve for the purpose of reading off the absorbance or transmittance value and the wavelength. The data associated with each data point can also be shown as a table.

6.6.1 Set up the wavelength scan

Press Wavelength Scan in the Main Menu. Press Options for

Parameter Setup.

 

Table 14 Wavelength scan setup options

Option

Description

 

 

More

For further Options

 

 

Store icon

To store the scan data

 

 

 

From the displayed list of stored scans, a record is selected for use as a reference

 

scan/superimposed scan. This can be highlighted or shown in the background in comparison

Reference Off/On

with the actual measured scan.

 

Note: This option is only available when there are stored scans with the same wavelength range

 

and step.

 

 

λ

To enter the wavelength spectrum and the scan interval

 

 

 

Enables the user to switch the display back and forth between the scan data tables

Select View

(wavelength/absorbance) and the graph of the curve.

 

Note: Select View will be activated after the first reading.

Cursor Mode

To select Track or Peak/Valley. The selection for this menu item determines to which points on

the graph the cursor moves.

 

Send Data

To send Data to a printer, computer or USB memory stick (Type A)

 

 

Integral: On/Off

The integral gives the area and the derivative of the integral gives the original function

 

 

94

Image 94
Contents January 2008, Edition Page Table of Contents Table of Contents 101 Table of Contents Section Specifications Specifications General Information Safety informationPrecautionary labels Use of hazard informationSource Lamp Safety Chemical and Biological SafetyGeneral Information Class 1 Laser Overview of product General InformationGeneral Information Unpack the instrument Section InstallationOperating environment Installation ConnectionCell compartment 1 100 mm adapter optional Multi Cell Holder Top and BottomInstallation Installation Multi Cell Holder Installation Multi Cell HolderBeam path Beam pathStart Up Power the instrument on and offChanging the language setting Language selectionStart Up Tips for the use of the touch screen Standard OperationsOverview Use of the alphanumeric keypadStandard Operations Main Menu Main Menu optionsStandard Operations Instrument Setup modeOperator ID Standard Operations Sample ID Date and time Display and sound preferencesLamp Control VIS-Lamp and UV-LampEffect caused by the lamp settings on the Measuring Mode Standard Operations Standard Operations PC and printer Printer setupUSB connector KeyboardPrinter Setup Print data Hach Data TransPassword Password deactivation Standard Operations Auto/manual data storage Store, recall, send and delete dataData Log Recall stored data from the data logSend data from the data log To send measurement data to a PCDelete stored data from the data log Recall stored data from wavelength scan or time course Data storage from wavelength scan or time courseSelect Wavelength Scan or Time Course to recall data Send data from wavelength scan or time course Option Press Wavelength Scan or Time Course and then Delete stored data from wavelength scan or time course Stored program options Stored ProgramsSelect a saved test/method enter user-specific basic data Stored programs optionsStandard Adjust Timer iconReading Mode Options Description Store Off/OnStandard Operations Use of program timers Set the reading modeTake single wavelength measurements continuous readings Set the dilution factorTake single wavelength measurements single reading Press OptionsMore…Dilution FactorRun a standard adjust Set the chemical form Change of the default setting of the chemical formPress Save as Default Press OptionsMoreChemical FormMeasurement/analysis of a reagent blank Standard Operations Run a reagent blankPress OptionsMoreReagent Blank Standard Operations Analysis of samples Add stored programs to the favorite programs listBarcode Programs Cell compartment #1Select the measuring range Changing the standard settingStandard Operations Complete a barcode 13 mm test/vial Select the chemical evaluation formBasic test-specific and sample-specific data settings Barcode program optionsOptions Description Send Data icon Sample blankEdit Program update options Update/edit barcode testsManual update of a barcode test Enter the NameEnter the Version Enter the Measurement ProcessTimer 1, Timer 2, Timer 3 and Timer Option 2 Manual check/revision of test data Update an existing barcode testOption 1 Manual update of test data Press OptionsMoreEditUpgrade of the instrument software Program a new testPress OK to return to the System Checks menu Standard Operations User Programs Advanced OperationsProgram a user method Program Name Advanced OperationsProgram Number Program TypeUnits Single wavelength settingsConcentration resolution number of decimal places Wavelength single wavelength program typeChemical form Multi wavelength settingsAbsorbance formula multi-wavelength program type Wavelength λx List of available absorbance formulasConcentration Factor Kx Calibration settings Calibration settings for single and multi wavelength modeCalibration by entering calibration values Advanced Operations Advanced Operations Advanced Operations Additional user-defined parameters and functions Store a user programUpper and lower limits of the measuring range Timer 1 / Timer 2 / Timer 3 / Timer Chemical form 2 / chemical form 3 / chemical formMeasurement process Advanced Operations Free programming program typeDefinitions of the program points Content and definition of the keys Key / Zeroing Enter a new element of a measuring sequenceProcess Timer key Key / ReadingInsert an element of a measuring sequence Enter the calibration formula evaluation formulaDelete an element of a measuring sequence Enter calibration formula C1Screen Key Description Edit formula key descriptionsNew Number Enter the next calibration formula C2 or Cn C2 OnSave a free programming user program Enter variablesSelect Conc. key Advanced Operations Select a user program 4 Add, edit and delete user programs from the favorites listEdit Add to FavoritesDelete Recall a favorite program Favorite ProgramsDelete a favorite program Standard Addition monitoring/checking results Measures to identify other interference factors ChecklistMethod Check the standard solution Peak volume Standard additionSample volume Complete a standard additionPeak volume/sample volume methods Press OptionsMoreStandard AdditionDescription of the table of readings Advanced Operations Single wavelength setup options Set up single wavelength modeProgram Concentration factor Concentration resolutionOptions Description Take single wavelength measurements single reading Reading modeTake single wavelength measurements continuous readings Set the reading mode at different wavelengths Multi-wavelength setup optionsAbsorbance formula Press Absorbance Formula Concentration factor For data storage, see .3.1.1 on Set up the wavelength scan Wavelength scan setup optionsOption Description Setting wavelengthScale & Units Select view displaying table Cursor modeIntegral Wavelength scan reading Scale & units Press Scale & UnitsNavigating the wavelength scan Advanced Operations Work with reference scans Highlight Reference100 Time course of absorbance/transmittance Time course setup parametersTime course setup options 101Scale & units Time course scan readingTime & interval 102Analysis of time course data 103Navigating the time scan Navigation of a time scan or a time scan analysis104 Instrument information System checks105 Optical checks Advanced Operations Upgrade of the instrument software106 Wavelength check 107Noise check 108109 Stray light check 110Absorbance check 111Drift check 112Verification kit Press Nominal Values 113Press Verification 114Lamp history Output checks115 Service time Advanced Operations Factory service116 Instrument Backup 117Restore backup data 118119 120 Spectrophotometer MaintenanceCleaning requirements DisplayTungsten lamp replacement Lamp replacementMaintenance 122Deuterium lamp UV replacement 123Fuse replacement 124Filter pad replacement Filter pad maintenance125 Base of DR 5000 with filter grid 126Section Troubleshooting 127Troubleshooting 128129 Section Replacement PartsDescription Cat.-No Replacement Parts 130Contact Information 131Contact Information 132Limitation of Remedies Limited WarrantyLimitations 133Limited Warranty 134Index 135136