Energizer ATX612 specifications Fail, Pump, Fault, R O E S S, 02 CAL

Page 9

BATTERY

 

4.6.6 BATTERY FAILURE

 

When the battery has insufficient charge to operate the

FAIL

 

instrument, “BATTERY FAIL” is displayed. The visual

 

 

alarm will be activated and the instrument will emit a

 

 

short beep once every second for approximately 30

 

 

seconds after which the instrument will turn itself off.

 

 

Recharge or replace the battery (See Section 8.2).

 

4.6.7 FAULT INDICATION

 

The ATX612 will emit a short beep once a second if a

 

newly installed sensor does not agree with the previous

 

sensor type for that position. The corresponding display

 

position will be blank. This fault indication also occurs

 

when a sensor becomes disconnected or a combustible

 

sensor fault is detected during normal operation. Installed

 

sensor types are accepted and displayed only after a

 

successful calibration has been completed.

 

The sampling pump system is equipped with a low flow

 

detection alarm. A low flow alarm condition will occur if

PUMP

the dust filter/water stop becomes clogged or the sample

line becomes blocked in any way. The instrument will

FAULT

sound a continuous tone, the red alarm light bar will flash

 

and the display will read PUMP FAULT. If this should

 

occur, replace the dust filter/water stop immediately (See

 

Section 9). The unit will not operate and a low flow alarm

 

condition will remain if the dust filter/water stop is

 

removed from the instrument. The PUMP FAULT alarm

 

will be reset once the obstruction has been cleared from

 

the sample line.

5. CALIBRATING THE ATX612

The ATX612 is a potential life saving device. Recognizing this fact, Industrial Scientific Corporation recommends that a functional (“bump”) test be performed on every instrument prior to each use. A functional test is defined as a brief exposure of the monitor to a known concentration of gas(es) for the purpose of verifying sensor and alarm operation. It is not intended to be a measure of the accuracy of the instrument.

Industrial Scientific also recommends that a full instrument calibration be performed using a certified concentration(s) of calibration gas(es) monthly to ensure maximum accuracy.

12

If an instrument fails to operate properly following any functional “bump” test, a full instrument calibration should be performed prior to use.

Calibration is most accurate when the instrument has been in a stable temperature environment for at least one hour prior to calibrating.

NOTE: Instrument zero and oxygen span calibration should be performed in clean air containing 20.95% (21%) oxygen. If you are measuring a known combustible gas, use a known concentration of that gas for calibration. For general combustible gas measurement, Industrial Scientific Corporation recommends calibrating to pentane in the 15-50% LEL range. The measured LEL concentration of gases other than the calibration gas may not correspond on a one-to- one basis with the monitor reading. Always use teflon or teflon-lined tubing when calibrating.

The ATX612 utilizes a one-button calibration system. When using multi-gas cylinders, a full instrument calibration can be performed in a single step. Multi-gas cylinders are available for the most common instrument configurations.

To calibrate the ATX612:

From the normal READING mode, press the MODE switch twice to access the ZERO operating mode.

Press the (E) key to start the instrument zeroing process. The instrument display will indicate ZEROING.

When zeroing is complete the instrument display will indicate O2 CAL and will show the current full span value of the oxygen sensor. The oxygen sensor will be calibrated to 21.0 in approximately 30 seconds.

Instrument zeroing and oxygen calibration may be aborted at any time by pressing the MODE switch.

NOTE: Zeroing the instrument in clean air is preferred, provided that there is no trace of toxic or combustible gas. If the air purity is uncertain, use a cylinder of zero grade air to zero the instrument and span the oxygen sensor.

At the completion of the oxygen sensor calibration, the instrument will emit a short beep and the display will indicate GO CAL. The message PRESS (E) to CAL will scroll across the bottom of the display. Press (E) to continue calibrating the remaining sensors. If you ignore this message the instrument will return to the normal operating mode in approximately eight seconds.

Z E R O

P R E S S

Z E R O I N G

2 1 . 0

02 CAL

13

Image 9
Contents ATX612 OUR Mission Page ATX612 Features Charging the Battery Nicad Battery Option Unpacking the InstrumentInstrument Operation Fast charged, in trickle CO H2S Hold ReleasePPM ZeroPK CLR Date 5/21W a T E L Fail FaultPump R O E S SCode Changing Instrument SettingsSetspan SetspanLEL AlarmsLatch Instant0 6 4 0 0 0 Replacing Alkaline Battery Cells MaintenanceCombustible Sensor REMOVAL/ Replacement TOXIC/OXYGEN Sensor REMOVAL/ ReplacementReplacement Parts Changing the Internal FilterPage Specifications Default Alarm Settings GAS LOW Alarm High AlarmATX612 Options & Ordering Information Limitation of Liability Warranty