28 | ST60 Tridata Instrument Owner’s Manual |
•Lock the shallow alarm.
•Set the required units for speed readings.
•Set the speed resolution.
•Select the log distance units
•Either calibrate the speed reading to Speed Over Ground (SOG) or manually apply a calibration factor, to obtain correct speed through the water.
•Select temperature units.
•Calibrate for correct temperature readings.
•Set timer alarm buzzer on or off.
Separate routines are provided for the User calibration of the depth and speed functions. To carry out either of these routines:
1. Power up the ST60 Tridata instrument.
To start User calibration
hold down
depth and speed
for approximately 2 seconds
User calibration entry screen
press either | depth | see User calibration - depth | illustration |
or | speed | see User calibration - speed | illustration |
Starting User calibration |
|
|
2.Press the depth and speed keys for approximately 2 seconds so that the User calibration entry screen is displayed.
3.Carry out the User calibration procedures for Depth and Speed.