Raymarine ST60 Tridata manual Shallow alarm lock

Page 42

30

ST60 Tridata Instrument Owner’s Manual

depth from the water line. In order to do this, you need to know the vertical separation between the transducer position and:

The bottom of the keel.

The water line.

Use the trip (decrement) and reset (increment) keys to set the required offset value:

If you want to display the depth reading from the transducer, set a value of 0.0.

If you want to apply a water line offset, adjust the displayed reading until the appropriate positive offset value is shown.

If you want to apply a keel offset, adjust the displayed reading until the appropriate negative offset value is shown.

+ve offset

Offset value

-ve offset

values

of 0.0

values

Depth offsets

 

D4352-2

Shallow alarm lock

When set to on, prevents alteration to the shallow depth alarm threshold.

Image 42
Contents ST60 Tridata Instrument Page EMC conformance Safety noticesImportant information Handbook informationPage Contents Chapter Page Page Introduction Data inputsSeaTalk Remote control Mounting optionsStand alone operation Parts supplied Tridata Normal operation OperationGetting started Displayed informationDepth Current depth displaySpeed Depth alarm threshold displaysAdjusting alarm thresholds Boat speedMaximum speed Average speedVelocity made good to windward Trip LogRace-start timers Water temperatureTimers Trip screenAlarms Display settingsIllumination Contrast Servicing and safety Maintenance and FaultfindingMaintenance InstrumentFixing faults Fault findingPreliminary procedures CablingTechnical support World wide webTelephone help line Help us to help you Site requirements InstallationPlanning your installation TransducersTransducer siting Instrument EMC Installation Guidelines ST60 instrument dimensionsSuppression Ferrites Connections to Other EquipmentFitting the instrument ProceduresUnpacking Surface mountingFlush mounting Surface mountingFitting the low-profile bezel Flush mounting procedure Bracket mounting Fitting transducer Running transducer cableConnecting the instrument Signal connectionsTypes of connection Power supply connections SeaTalk systemsStand alone instruments User calibration Calibration Introduction Speed readingsStarting User calibration Depth units Depth offsetShallow alarm lock Set speed units Set speed resolutionSetting the correct speed Set log unitsUser calibration speed Temperature calibration Adjust to SOGSet temperature units Timer alarm buzzerIntermediate calibration Speed calibration Page Calibration From Speed Leaving Intermediate calibration Response settingsDealer calibration User calibration on/offDealer calibration Boat show mode Factory defaultsLeaving Dealer calibration Page TOP Page Shaded area to be removed