Toshiba TIC-LF494B manual Ordering Information, Calibration Range, Standard Flow Range

Page 11

TIC-LF494B

Calibration Range

If the calibration range is not specified, the standard range as shown below will be used. If the range is specified, we will use the specified range for calibration.

Table 4. Standard Flow Range

Meter size

Standard flow range

inch(mm)

Flow rate

Flow rate

Flow velocity

(gal/min)

(m³/h)

(m/s)

 

1 (25)

75

6

3.395

1 1/2 (40)

175

15

3.316

2 (50)

300

25

3.537

3 (80)

650

60

3.316

4 (100)

1,000

100

3.537

Note: The unit of "gal/min" is not exchanged (converted) by "m3/h".

About establishment environment

Do not store or install the flowmeter :

Where there is direct sunlight.

Where excessive vibration or mechanical shock occurs.

Where high temperature or high humidity conditions exist.

Where corrosive atmospheres exist.

Places that can be submerged under water.

Where there is a sloped floor. To put the flowmeter temporarily on the floor, place it carefully with something, such as a block, to support it so that the flowmeter will not topple over.

In areas like the following, there may be the case that infrared switches do not function correctly. (If these are unavoidable, use an appropriate cover.)

(1)Where unit (operation panel) is exposed to direct sunlight, reflection of light onto window pane and diffused light reflection.

(2)Where smoke and steam may occur.

(3)Where exposed to direct snow, ice or mud.

Ordering Information

1.When ordering the LF490 series flowmeters, refer to Tables 5 and 6 (Type Specification Codes). An entry must be made for each of the columns in each of these tables.

2.Fluid characteristics:

(1)Type of fluid to be measured and its characteristics

(2)Fluid temperature

(3)Fluid pressure

(4)Electrical conductivity of the fluid

3.Measuring range

4.I/O function setting

5.Ordering scope:

Flow calibration data: (required or not)

6.Other items

Specifications other than standard items

Consult a Toshiba representative before ordering when choosing materials of the wetted parts such as lining, electrodes, and grounding rings.

11

Image 11
Contents Combined type Separate type IntroductionModel LF490 and LF494 Detectors SpecificationsOverall Specifications Model LF610 and LF612 converters Cable connection ports Cable glands Power supplyDimensions and Weights Surge protectionDimensions Combined type InstallationDimensions Separate type Ferrule dimensions Combined type LF490/LF610 and LF494/LF610F flowmeters External ConnectionsSeparate type LF490/LF612 and LF494/LF612F flowmeters To select the meter size Wiring PrecautionsWiring Precautions Profibus Meter SizeAbout establishment environment Ordering InformationCalibration Range Standard Flow RangeSpecification Code Sanitary type detector LF490 Series Type Specification Code for Maintenance Parts Specification Code for convertersThis Page Intentionally Blank TIC-LF494B ISO9001 and ISO14001 are certified