Toshiba TIC-LF494B manual Model LF610 and LF612 converters

Page 3

TIC-LF494B

Model LF610 and LF612 converters

Input signals

Analog signal — the voltage signal from detector, proportional to process flow rate (for LF612 separate type converter).

Digital input DI (opt.)

Signal type: 20 to 30Vdc voltage signal

Input resistance: 2.7kΩ

Number of inputs: one point

DI function — One of the following functions can be assigned to the optional DI signal.

Range switching — Selects either the higher or lower range in the unidirectional or bidirectional 2-range setting.

Totalizer control — Starts and stops the built-in totalizer.

Fixed-value outputs —Outputs fixed-values for current and pulse outputs.

Zero adjustment — Executes zero adjustment (on-stream at zero flow rate).

Output signals

Current output:

4–20mAdc (load resistance 0 to 750Ω) Note: The current output cannot be used with

the PROFIBUS-PA ccommunication. Digital outputs — One point (std.) and one more

point is optionally available as follows.

Digital output DO1 (std.):

Output type: Transistor open collector

Number of outputs: One point

Output capacity: 30Vdc, 200mA maximum

Digital output DO2 (opt.):

Output type: Solidstate relay output (non polarity)

Number of outputs: One point

Output capacity: 150Vdc, 150mA maximum or 150Vac (peak to peak), 100mA maximum

DO1 and DO2 functions — One of the following functions can be assigned to DO1 (std.) and/or DO2 (opt.)

Pulse output (available only for DO1,DO2) Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1)

3.6 to 360,000 pulses/hr (DO2) (Over 3,600,000 pulses/hr, auto-setting)

Pulse width: 0.5 to 500ms (but less than half of the period for 100% flow rate)

Note: The same and simultaneous pulse is not available between DO1 and DO2.)

Multi-range selection outputs (Note 1)

High, High high, Low, and/or Low low alarm outputs (Note 2)

Empty pipe alarm output

Digital Output Active Status (DO1 and DO2) (Note 2)

Preset count output

Converter failure alarm output

Note 1: Two outputs (DO1 and DO2) are needed for 4-range switching and forward/reverse 2-range switching.

Note 2: Normal Open (default set) or Normal Close is selected for alarm outputs when programming.

When power failure occurs, unit will be fault to

Normal Open.

Communications output:

HART (std.) — Digital signal is superimposed on 4–20mAdc current signal as follows:

Conforms to HART protocol

Load resistance: 240 to 750Ω

Load capacitance: 0.25µF maximum

Load inductance: 4mH maximum

PROFIBUS (opt.)

Protocol : PROFIBUS-PA

Baud rate : 31.25kbps

Bus voltage : 9-30VDC

Consumption electric current of bus:less than 16mA

Manufacture Ident-No. : 093BHEX

Standard Ident-No. : 9740HEX

Slave address : 0-126 (Default address is 126)

Profile : Profile Ver.3.01 for Process Control Devices

Function blocks : AI(Flow)×1 , Totalizer×1

LCD display: Full dot-matrix 128×128 dot LCD display (back-light provided)

The data on the LCD inside the converter can rotate to 90, 180, and 270 degrees by a software, without rotating the indicator itself. (Combined type only)

Parameter settings — Parameters can be set as follows:

IR Switches: Three key switches are provided to set configuration parameters.

Digital communication: The AF900 hand-held terminal or PROFIBUS is needed to set parameters.

•Zero adjustment: Zero point adjustment can be started by pressing the switch in the converter.

•Damping: 0.5 to 60 seconds (selectable in one second increments)

“Field re-verification” Mag-Prover– Toshiba’s

Zero span calibration tool allows unit to be re-calibrated and verified using an internal software program. (For more information contact Toshiba International Corp.)

Conditions when power fails:

Parameter setting values are stored in non-volatile memory and the values will be restored when the power returns to normal condition. The outputs and display will remain as follows when power fails.

• Current output: 0mAdc

3

Image 3
Contents Combined type Separate type IntroductionSpecifications Overall SpecificationsModel LF490 and LF494 Detectors Model LF610 and LF612 converters Cable connection ports Cable glands Power supplyDimensions and Weights Surge protectionDimensions Combined type InstallationDimensions Separate type Ferrule dimensions Combined type LF490/LF610 and LF494/LF610F flowmeters External ConnectionsSeparate type LF490/LF612 and LF494/LF612F flowmeters To select the meter size Wiring PrecautionsWiring Precautions Profibus Meter SizeAbout establishment environment Ordering InformationCalibration Range Standard Flow RangeSpecification Code Sanitary type detector LF490 Series Type Specification Code for Maintenance Parts Specification Code for convertersThis Page Intentionally Blank TIC-LF494B ISO9001 and ISO14001 are certified