Servicing the Interface Module
Verify the Interface Module Performance
If the "Ethernet" selection appears in the Communications menu, then go to step 5.
If "Ethernet" does not appear in the Communications menu, perform the following steps.
aRemove power from the HP 16500B, disconnect the power cable, then remove the top cover.
bRemove and inspect the interface module cable for signs of damage. If the cable appears defective, replace it.
cEnsure that the cable is connected according to the installation procedures and that all cables are properly seated. For installation procedures, refer to chapter 2, "Installing the Interface Module."
dInstall the top cover of the HP 16500B mainframe, connect the power cable, and reapply power.
If Ethernet still does not appear, then suspect a defective CPU board. Refer to the HP 16500B/16501A Service Guide for infomation on testing and replacing the CPU baord.
5Touch Configuration, then touch Test. When the test menu appears, touch the blue field to load the performance verification test system.