Emerson 5081T-FF/FI instruction sheet Physical Specifications, Transmitter Specifications @ 25C

Page 2

MODEL 5081T-FF/FI

SPECIFICATIONS

PHYSICAL SPECIFICATIONS

HOUSING: Epoxy-polyester painted over low-copper alu- minum. Neoprene O-rings on cover. 160.5 mm x 175.3

mmx 161.3 mm (6.3 in. x 6.9 in. x 6.4 in.) DIAMETER: 155.4 mm (6.1 in.)

ELECTRICAL CONDUIT OPENINGS: 3/4 in. FNPT

POWER SUPPLY AND LOAD: A power supply voltage of 9 Vdc to 32 Vdc at 22 mA is required; Intrinsically Safe installations may be limited to a maximum of 2-3 transmitters per node, depending on the barrier used.

LOCAL READOUT:

Main Display is 4 digits, 20 mm tall (0.8 in.) Message Display is ten digits, 7 mm tall (0.3 in.)

AUTOMATIC TEMPERATURE COMPENSATION:

3-wire Pt100 RTD

Conductivity: 0 to 200 °C (32 to 392 °F)

%concentration: 0 to 100 °C (32 to 212°F) AMBIENT TEMPERATURE: -20 to 65° C (-4 to 149° F)

RELATIVE HUMIDITY: 0-95% with enclosure sealed.

CE: EMI/RFI certified: EN61326-1

HAZARDOUS AREA CLASSIFICATION:

Intrinsic Safety:

Class I, II, III, Div. 1

Groups A-G

T4 Tamb = 70°C

Exia Entity

Class I, Groups A-D

Class II, Groups E-G

Class III

T4 Tamb = 70°C

ATEX 0600

II1 G Baseefa03ATEX0399 EEx ia IIC T4 Tamb = -20°C to +65°C

Non-Incendive:

Class I, Div. 2, Groups A-D

Dust Ignition Proof

Class II & III, Div. 1, Groups E-G

NEMA 4X Enclosure

Class I, Div. 2, Groups A-D

Suitable for Class II, Div. 2, Groups E-G

T4 Tamb = 70°C

Explosion-Proof:

Class I, Div. 1, Groups B-D

Class II, Div. 1, Groups E-G

Class III, Div. 1

Class I, Groups B-D

Class II, Groups E-G

Class III

Tamb = 65°C max

TRANSMITTER SPECIFICATIONS @ 25°C

MEASURED RANGE*: 50 to 2,000,000 µS/cm (see chart)

ACCURACY: ± 1.0% of reading

REPEATABILITY: ± 0.25% of reading

STABILITY: 0.25% of output range/month, non-cumulative

AMBIENT TEMPERATURE COEFFICIENT: ± 0.2% of FS/°C

COMPATIBLE RTD: 100with Automatic Recognition

TEMPERATURE SLOPE ADJUSTMENT: 0-5%/° C

%CONCENTRATION RANGES: Sodium Hydroxide: 0 to 15% Hydrochloric Acid: 0 to 16%

Sulfuric Acid: 0 to 25% and 96 to 99.7%

LOOP SPECIFICATIONS

LOOP ACCURACY: With a standard Model 228 or 225 sensor with 20' cable, laboratory accuracy at 25°C can be as good as ±2% of reading and ±50 µS/cm.

To achieve optimum performance, standardize the sen- sor in the process at the conductivity and temperature of interest.

Results under real process conditions, at different tem- peratures, or using other sensors may differ from above.

RTD ACCURACY: Utilizing a perfect 100 Ohm RTD after 1 point temperature standardiztion, temperature read- ing can be as good as ±0.5°C.

RECOMMENDED SENSORS:

Model 222 Flow-Through

Model 225 Clean-In-Place (CIP)

Model 226 Submersion/Insertion

Model 228 Submersion/Insertion/Retractable

Model 242 Flow-Through*

*Model 242-06 or 242-08 with 5081T do not have Intrinsically Safe approvals.

RECOMMENDED RANGES FOR TOROIDAL SENSORS

 

Conductivity Sensor

 

 

 

 

 

 

 

Model Number

226

228

225

222 (1in.)

222 (2 in.)

242

 

 

 

 

 

 

 

 

 

Nominal Cell Constant

1.0

3.0

3.0

6.0

4.0

*

 

Minimum Conductivity (S/cm)

50

200

200

500

500

100*

 

 

 

 

 

 

 

 

 

Maximum Conductivity (S/cm)

1,000,000

2,000,000

2,000,000

2,000,000

2,000,000

2,000,000*

 

 

 

 

 

 

 

 

* Model 242 values depend on sensor configuration and wiring.

NOTE: Values shown are for 25°C conductivity with a temperature slope of 2% per degree C. The maximum range

2

value will be lower for solutions with a higher temperature slope. Minimum conductivity depends on sensor.

 

Image 2
Contents Essential Instructions Read this page Before ProceedingLoop Specifications Physical SpecificationsTransmitter Specifications @ 25C Recommended SensorsInfrared Remote Controller Figure I T N E X T25.0C 12.00mA Process DisplayDiagnose Menu Program MenuCalibrate Menu Wiring Wiring Model 5081T-FF/FIWiring Through a Junction BOX Power WiringUnpacking and Inspection InstallationInstallation Rotating the DisplayPage Page Page Page Page FM REV a 2X Full RModel Finishsilkscreen Black Epoxy Paint Baked1400298 Wiring1400298 a LabelHazardous Classified Locations 14002841400284 NON-HAZARDOUS LocationsCSA 140 125125 650 ±.015 1400285 1400285Hazardous Area Unclassified AreaBaseefa REV a II 1 G 1180125 4X R Artwork is Sheet 2 No Change Without Baseefa ApprovalFisco Overview MaintenanceTransmitter Maintenance Exploded View of Model 5081-T-FF/FI TransmitterDescription QUICK-START Engineering SpecificationsTo reset transmiter to factory default settings Specifications subject to change without notice