Iomega EP510 manual Series I/P and E/P Transducer Specifications

Page 10

AM-4109/1104 IP510, EP510, IP511, EP511 SERIES

510 and 511 Series I/P AND E/P TRANSDUCER SPECIFICATIONS

Accuracy:

± 0.10% of output span, typical

± 0.25% of output span, maximum

 

Hysteresis:

0.01% of output span, typical

0.10% of output span, maximum

 

Dead Band:

Not measurable

 

 

Repeatability:

0.01% of output span, typical

0.10% of output span, maximum

 

Ambient Temperature Effect

± 0.004% of nominal span per ˚F, typical

± 0.022% of nominal span per ˚F, maximum

Zero:

 

Span:

± 0.013% of calibrated span per ˚F, typical

± 0.022% of calibrated span per ˚F, maximum

 

Compensated Temperature Range:

+32˚F to +122˚F

 

 

Operating Temperature Range

-20˚F to 160˚F

Buna-N elastomers:

0˚F to 160˚F

Viton elastomers:

 

 

 

Storage Temperature Range

-40˚F to 200˚F

Buna-N elastomers:

-15˚F to 200˚F

Viton elastomers:

 

 

 

Vibration Effect:

Less than 0.5% of SPAN per 1G,

5-2000 Hz, 3G maximum, 3 axes

 

Mounting Position Effect:

Not measurable

 

 

Loop Load, I/P transducer:

Less than 10 VDC drop at 20 mA

Less than 12 VDC drop at 50 mA

 

Supply Voltage, E/P Transducer

9 VDC to 28 VDC, less than 20 mA

Intrinsically Safe/Nonincendive:

9 VDC to 40 VDC, less than 20 mA

General Purpose:

 

 

 

Supply Voltage Effect:

Not measurable

 

 

Signal Impedance, E/P Transducer:

6000 Ohm minimum

 

 

RFI/EMI Effect (NEMA 4X):

Less than 0.25% of SPAN, 10V/meter, 20-1000 MHz.

 

 

Supply Pressure Requirements:

See TABLE 1

 

 

Supply Pressure Effect:

Not measurable

 

 

Air Consumption:

6.0 SCFH maximum

 

 

Forward Flow Capacity (Supply):

Standard Range 7 SCFM

Extended Range (over 30 PSI) 12 SCFM

 

Port Sizes:

Pneumatic: 1/4" NPT

Electrical: 1/2" NPT

 

 

 

10

Image 10
Contents EP510, IP510, EP511, IP511 Series Servicing North America Output ContentsAIR Connections Supply GaugeMounting General Direct MountingTerminal Block Transducer Models DirectionsElectrical Connections CalibrationElectrical Maintenance MaintenancePneumatic Maintenance Nonincendive Entity ParametersEnclosures Explosion & DUST-IGNITION Proof Certifications Nema 4X, Replacement Cover O-RINGSNema 4X, Replacement Parts for Cover Locking Device Part Name Part NumberKits KIT Name511 Series I/P and E/P Transducer Specifications WARRANTY/DISCLAIMER Temperature