SAFETY CHECK-OUT
LEAKAGE TEST
The AC leakage from any exposed metal part to earth ground and from all exposed metal parts to any exposed metal part having a return to chassis, must not exceed 0.5 mA(500 microamperes). Leakage current can be measured by any one of three methods.
1.A commercial leakage tester, such as the Simpson 229 or RCA
2.A
3.Measuring the voltage drop across a resistor by means of a VOM or battery- operated AC voltmeter. The “limit” indication is 0.75 V, so analog meters must have an accurate low voltage scale.
The Simpson’s 250 and Sanwa
HOW TO FIND A GOOD EARTH GROUND
Trouble Light
AC Outlet Box
Ohmmeter
A
If the retaining screw is to be used as your earth ground, verify that it is at ground by measuring the resistance between it and a
If a
Figure B. Checking for earth ground.