Guidance and manufacturer’s
for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer’s
The ET2400LM is intended for use in the electromagnetic environment specified below. The customer or the user
of the ET2400LM should assure that it is used in such an environment.
Immunity test | IEC 60601 | Compliance level | Electromagnetic environment- | |
test level |
|
|
| guidelines |
|
|
|
|
|
Electrostatic | ± | 6 kV contact | ± 6 kV contact | Floors should be wood, concrete or |
discharge(ESD) | ± | 8 kV air | ± 8 kV air | ceramic tile. If floors are covered with |
IEC |
|
|
| synthetic material, the relative humidity |
|
|
|
| should be at least 30%. |
|
|
|
| |
Electrical Fast | ± 2 kV for power | ± 2 kV for power | Mains power quality should be that of a | |
transient/burst | supply lines | supply lines | typical commerical or hospital | |
|
|
|
| environment. |
IEC | ± 1 kV for input/output | ± 1 kV for input/output |
| |
| lines | lines |
| |
|
|
|
| |
Surge | ± 1 kV line(s) to line(s) | ± 1 kV line(s) to line(s) | Mains power quality should be that of a | |
IEC | ± 2 kV line(s) to earth | ± 2 kV line(s) to earth | typical commerical or hospital | |
|
|
|
| environment. |
|
|
|
| |
Voltage dips, short | <5% UT | <5% UT | Mains power quality be that of a | |
interruption and | (>95% dip in UT) | (>95% dip in UT) | typical commerical or hospital | |
voltage variations | for 0.5 cycle | for 0.5 cycle | environment. If the user of the ET2400LM | |
on power supply |
|
|
| requires continued operation during power |
input lines | 40% UT | 40% UT | mains interruptions, it is recommended that | |
| (60% dip in UT) | (60 % dip in UT) | the ET2400LM be powered from an | |
IEC | for 5 cycles | for 5 cycles | uninterruptible power supply or a battery. | |
| 70% UT | 70% UT |
| |
| (30% dip in UT) | (30% dip in UT) |
| |
| for 25 cycles | for 25 cycles |
| |
| <5% UT | <5% UT |
| |
| (>95% dip in UT) | (>95% dip in UT) |
| |
| for 25 cycles | for 25 cycles |
| |
|
|
|
| |
Power frequency | 3 A/m | 3A/m | Power frequency magnetic fields | |
(50/60 Hz) |
|
|
| should be at levels characteristic of a |
magnetic field |
|
|
| typical location in a typical commerical |
|
|
|
| or hospital environment. |
IEC |
|
|
|
|
|
|
|
| |
NOTE UT is the a.c. mains voltage prior to application of the test level. |
|
x