APPENDIX A: EMC INFORMATION
Guidance and Manufacturer’s Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
Emissions Test |
| Compliance |
| Electromagnetic |
Harmonic emissions |
| Class A |
| ||
|
| |
Voltage fluctuations | / |
|
Flicker emissions |
| Complies |
|
|
The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public
Guidance and Manufacturer’s Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.
|
|
|
|
|
|
|
| Electromagnetic |
| |
| Immunity Test |
| IEC60601 test level |
|
| Compliance | ||||
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
| ||
Electrostatic |
|
|
|
|
|
| Floors should be wood, concrete or | |||
| ±6kV contact |
|
| ±6kV contact |
| ceramic tile. If floors are covered | ||||
Discharge(ESD) |
|
|
|
| ||||||
| ±8kV air |
|
| ±8kV air |
| with synthetic material, the relative | ||||
|
|
|
|
|
| humidity should be at least 30 %. | ||||
|
|
|
|
|
|
|
| |||
|
|
|
|
| ||||||
Electrical fast |
| ±2kV for power supply | ±2kV for power supply | Mains power quality should be that | ||||||
| transient/ burst |
| line |
|
| line |
| of atypical commercial or hospital | ||
|
| ±1kV for input/out line |
|
| ±1kV for input/out line |
| environment |
|
| |
|
|
|
|
|
| |||||
Surge |
| ±1kV for differential | ±1kV for differential | Mains power quality should be that | ||||||
| mode |
|
| mode |
| of atypical commercial or hospital | ||||
|
|
|
| |||||||
| ±2kV for common mode | ±2kV for common mode | environment. |
|
| |||||
|
|
|
|
| ||||||
|
|
|
| |||||||
Voltage dips, | <5 % UT(>95 % dip in | <5 % UT(>95 % dip in | Mains power quality should be that | |||||||
short | UT)for 0,5 cycle |
|
| UT) for 0,5 cycle |
| of atypical commercial or hospital | ||||
interruptions and | 40 % UT(60 % dip | in | 40 % UT(60 % dip | in | environment. If the user of this | |||||
voltage | UT)for 5 cycles |
|
| UT) for 5 cycles |
| device requires continued operation | ||||
variations on | 70 % UT(30 % dip in | 70 % UT(30 % dip in | during power mains interruptions, it | |||||||
power supply | UT)for 25 cycles |
|
| UT) for 25 cycles |
| is recommended that the device be | ||||
input lines | <5 % UT(>95 % dip in | <5 % UT(>95 % dip in | powered from an | uninterruptible | ||||||
UT)for 5 sec |
|
| UT) for 5 sec |
| power supply or a battery. | |||||
|
|
|
|
|
|
|
|
| ||
Power frequency |
|
|
|
|
|
| Power frequency | magnetic fields | ||
(50/60Hz) |
| 3 A/m |
|
| 3 A/m |
| should be at levels characteristic of | |||
| magnetic field |
|
|
|
| atypical location in a typical | ||||
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
| commercial or hospital environment. | |||
|
|
|
|
|
|
|
|
|
|
|
NOTE: UT is the a.c. mains voltage prior to the application of the test level
| 21 |