Graham Field 765000 user manual INS-LAB-RevB10

Page 22

Guidance and Manufacturer’s Declaration- Electromagnetic Immunity:

This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment.

Immunity Test

IEC60601 test

level

Electromagnetic

Compliance Environment-Guidance

Conducted RF IEC 61000-4-6

Radiated RF IEC 61000-4-3

3Vrms150 kHz to 80 MHz outside ISM bandsa

3 V/m 80 MHz to

2.5 GHz

3Vrms

3V/m

Portable and mobile RF

communications equipment should be used no closer to any part of this device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

d = 1.2 P 150kHz to 80MHz

d = 1.2 P 150kHz to 80MHz d = 2.3 P 80 MHz to 2.5G MHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).b

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey c, should be less than the compliance level in each frequency ranged.

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

22

765000-INS-LAB-RevB10

Image 22
Contents 765000 Contents Introduction Intended USE of this DeviceImportant Safety Precautions INS-LAB-RevB10 INS-LAB-RevB10 Equipment Symbols Pump Front Product DescriptionPump and Mattress System Pump RearPump Front Panel Front panel, first rowFront panel, second row Front panel, third row Power failure. Depress the alarm mute button to disable Pump and Mattress Setup InstallationUnpacking INS-LAB-RevB10 Getting Started OperationQuick Disconnect Transport CAP CPR FunctionCleaning MaintenanceGeneral Maintenance Air filter replacement Fuse replacementStorage Specifications Pump SpecificationLimited Warranty Appendix a EMC Information INS-LAB-RevB10 =1.2 P = 2.3 P GF Health Products, Inc