PUMP: FRONT PANEL

Front panel of pump is shown below. Descriptions of but- tons and indicators follow.

Auto-Adjust button;

Low pressure

Tuning button;

indicator

 

indicators at right of button

 

indicators at left of button

 

 

Therapy adjust button;

 

 

Maxfirm button

indicators at right of button

 

 

 

Panel lock button;

 

 

Service indicator

 

 

 

indicator at right of button

 

Power failure indicator

Power (ON/OFF) button;

Alarm mute button

indicator at right of button

 

 

 

Front panel, first row

Auto-Adjust button: The auto-adjust button adjusts the mattress pressure automatically based on the patient's weight. To remind the user that the auto-adjustprocess is ongoing, the auto-adjust indicator LEDs will periodically flash from left to right until the Ready LED illu- minates to show the process is complete. Three conditions trigger the auto-adjust function:

1.Press the auto-adjust button for two seconds; the mat- tress system automatically readjusts pressure setting.

2.When initial mattress inflation is complete, the mattress system automatically starts the auto-adjust function.

3.When the mattress system detects a significant change in patient's weight on the mattress (e.g. patient egress or ingress), the system automatically starts the auto- adjust function.

When auto-adjust is complete, the Ready indicator will illu- minate. To disable the auto-adjust function, press the auto- adjust button until auto-adjust indicators de-illuminate.

Note: During the auto-adjust operation, it is normal for the mattress system to cycle through a series of inflation and deflation.

765000-INS-LAB-RevB10

9

Image 9
Contents 765000 Contents Intended USE of this Device IntroductionImportant Safety Precautions INS-LAB-RevB10 INS-LAB-RevB10 Equipment Symbols Pump and Mattress System Product DescriptionPump Front Pump RearFront panel, first row Pump Front PanelFront panel, second row Front panel, third row Power failure. Depress the alarm mute button to disable Installation Pump and Mattress SetupUnpacking INS-LAB-RevB10 Operation Getting StartedQuick Disconnect CPR Function Transport CAPMaintenance CleaningGeneral Maintenance Fuse replacement Air filter replacementStorage Pump Specification SpecificationsLimited Warranty Appendix a EMC Information INS-LAB-RevB10 =1.2 P = 2.3 P GF Health Products, Inc