Graham Field 750000 user manual Operation, Pressure Adjustment

Page 14

OPERATION

1. Turn the power switch to the ON position.

2. Adjust the pressure adjust knob to the maximum position.

Info: Set the pressure adjust knob to the maximum position for the quickest inflation at set-up; the mattress can then be adjusted to the desired pressure.

3.The pump will begin delivering air into the

mattress. While the

mattress is filling, the low pressure indicator (yellow LED) will illuminate.

4. When the selected pressure level is

reached, the low pressure indicator will

de-illuminate, and the normal pressure indicator (green LED) will automatically illuminate.

5. Replace all sheets, blankets, and pillows on the bed.

PRESSURE ADJUSTMENT

Generally, a lighter patient will need a lower (softer) setting while a heavier patient will need a higher (firmer) setting, but pressure adjustment must ultimately be based on the patient's weight distribution. A short, stocky person may be the same weight as a tall, thin person, but their settings will not be the same. The mattress should be firm enough to support the patient, yet soft enough so the patient will sink down to eliminate pressure points. Test the pressure by placing four fingers below the mattress

14

750000-INS-LAB-RevC11

Image 14
Contents 750000 Contents Introduction Intended USE of this DeviceImportant Safety Precautions INS-LAB-RevC11 INS-LAB-RevC11 Equipment Symbols Product Description Pump and Mattress SystemPump rear Pump top panel INS-LAB-RevC11 Installation UnpackingPump and Mattress Installation INS-LAB-RevC11 Pressure Adjustment OperationCPR Function Quick DisconnectTransport CAP General Maintenance MaintenanceCleaning Fuse replacement StorageSpecifications Pump SpecificationWarranty 2011, GF Health Products, Inc