B&K 0-30V, 0-3A Maintenance, Fuse Replacement, Line Voltage Conversion, International Units

Models: 0-30V 0-3A

1 58
Download 58 pages 35.95 Kb
Page 23
Image 23
MAINTENANCE

MAINTENANCE

WARNING

The following instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing other than contained in the operating instructions unless you are qualified to do so.

FUSE REPLACEMENT

If the fuse blows, the CV or CC indicators will not light and the power supply will not operate. The fuse should not normally open unless a problem has developed in the unit. Try to determine and correct the cause of the blown fuse, then replace only with a fuse of the correct rating as listed below. The fuse is located on the rear panel (see Fig. 2).

OPERATION

FUSE VALUE

TYPE

120V

2.5A

SLOW BLOW

220/230/240V

1.25A

SLOW BLOW

 

Table 1. Fuse Values.

 

LINE VOLTAGE CONVERSION, INTERNATIONAL UNITS

The primary winding of the power transformer is tapped to permit operation from 120/220/230 or 240 VAC, 50/60 Hz line voltage. Conversion from one line voltage to another is done by a simple position change of the fuse receptacle underneath the line cord.

ADJUSTMENTS

This unit was accurately adjusted at the factory before shipment. Readjustment is recommended only if repairs have been made in a circuit affecting adjustment accuracy, or if you have a reason to believe the unit is out of adjustment. However, adjustments should be attempted only if a multimeter with an accuracy of ±0.5% DCV or better is available (B+K Precision Model 2703A or equivalent).

23

Page 23
Image 23
B&K 0-30V, 0-3A instruction manual Maintenance, Fuse Replacement, Line Voltage Conversion, International Units, Adjustments