B&K 0-30V, 0-3A Specifications, Constant Voltage Operation, Constant Current Operation, Metering

Models: 0-30V 0-3A

1 58
Download 58 pages 35.95 Kb
Page 7
Image 7
SPECIFICATIONS

SPECIFICATIONS

OUTPUT VOLTAGE:

0 to 30VDC, coarse and fine

adjustment.

 

OUTPUT CURRENT:

0 to 3A, coarse and fine adjustment.

CONSTANT VOLTAGE OPERATION

Voltage Regulation

 

Line (108-132 V):

0.01% + 3mV.

Load (no load to full load):

0.01% + 3mV.

Recovery Time:

100 µs typical.

Ripple Voltage

 

Peak-to-Peak:

2mV, typical

RMS:

1mV.

Temperature Coefficient

 

(0° to +35 °C):

300 PPm/°C

CONSTANT CURRENT OPERATION

Adjustable Current Limits:

5% to 100%

Current Regulation

 

Line (108-132 V):

0.2% + 3mA

Load

0.2% + 3mA

Current Ripple:

3mA typical.

METERING

 

Voltmeter

4 digit Green LED Display

Range:

0 to 99.99 V.

Accuracy:

(0.5% + 9 digits)*.

Ammeter

4 digit Red LED Display

Range:

0 to 9.999A.

Accuracy:

(0.5% + 9 digits)*.

(* = see note 1 at end of specifications)

POWER REQUIREMENTS:

Domestic: 120 VAC ±10%, 60 Hz

International: 120/220/230/240 VAC - 10%, 50/60 Hz

POWER CONSUMPTION:

Approximately 180W or less at

 

full load.

PROTECTION:

Reverse polarity protection,

 

Current limiting.

7

Page 7
Image 7
B&K 0-30V Specifications, Constant Voltage Operation, Constant Current Operation, Metering, Power Requirements, Protection