Philips 861304 owner manual Defibrillator says To stop all motion

Page 56

APPENDICES

48

defibrillator says:

... to stop all motion

... the shock was not delivered

... the shock button was not pressed

possible cause

The patient is being moved or jostled.

The environment is dry and movement around the patient is causing static electricity to interfere with ECG analysis.

Radio or electrical sources are interfering with ECG analysis.

The pads may not be making good contact with the patient’s skin.

The pads may be touching each other.

The pads may be damaged.

Shock has been advised but the shock button has not been pressed within 30 seconds.

recommended action

Stop CPR; do not touch the patient. Minimize patient motion. If the patient is being transported, stop the vehicle.

Responders and bystanders should minimize motion, particularly in dry environments that can generate static electricity.

Check for possible causes of radio and electrical interference and turn them off or remove them from the area.

Press the pads firmly to the patient's chest.

Make sure the adhesive pads are correctly positioned on the patient.

Replace the pads if necessary.

When next prompted, press the Shock button to deliver shock.

Philips Medical Systems

HEARTSTART FRx 861304 OWNER’S MANUAL

Image 56
Contents Edition 861304N E R ’ S M a N U a L Philips Medical Systems Quick Reference Intentionally blank HeartStart FRx About This Edition Edition History CopyrightAuthorized EU Representative Device TrackingContents Appendices Description Sudden Cardiac ArrestState and local requirements Indications for UseTraining and practice Package contents For more informationSetting up the FRx 1A. Remove Pads Case from packagingPhilips Medical Systems Recommended accessories Using the HeartStart FRx OverviewOpen the pads case as shown below. Peel off one pad Press the green on/off buttonFollow the FRx’s voice instructions Press the flashing orange Shock button if instructed If a shock is neededTreating infants and children If a shock is not neededKey if available When emergency medical services arrive Infant/Child KeyPhilips Medical Systems After using the HeartStart FRx After each useFRx data storage Routine maintenance Maintaining the HeartStart FRxDisposing of the FRx Cleaning the FRxPeriodic checks Troubleshooting tips Appendices Philips Medical Systems Accessories Appendices Glossary of terms HeartStart Event Review ButtonInfant/Child Key Coaching, if selected, appropriate for infants and childrenWhen a battery is inserted Patient care in the AED modeOn/Off button Pads See Smart PadsStandby mode Blinking green Ready lightConsciousness, absence of respiration, and lack of a pulse Waveform See Smart biphasic waveformShock is advised, as a reminder not to touch the patient Glossary of symbols/controlsStandby mode gives device status Symbol  #  Appendices Please read these warnings and precautions carefully Approved accessories are usedHazard. Do not sterilize the FRx or its accessories AccessoriesPrecautions HeartStart FRx 861304 Defibrillator specifications Technical informationPhysical EnvironmentalProtected Controls and indicatorsCategory Specifications Sealing Defibrillation waveform Current aDisarm AED mode Adult shock delivery vector Category Energy Charge control Shock-to-shock cycle timeEnergy dose Newborn 14 J/kgRhythm class Samplea size Performance Lower ECG analysis systemECG analysis performance Accessories specifications HeartStart Smart Pads IIInfant/Child Key M5070A battery and 989803139301 TSO C-142* batteryCategory Battery type Capacity Regulations Environmental considerationsEnvironment Configuration Overview Device optionsPatient treatment protocol options NSA pause type Settings Defibrillator conductsBackground monitoring. If a Is detected in a motionlessCPR prompt Settings CPR1 Instructs the user to begin CPRCPR reminder voice Instructs Instructions provided at User that it isWhen an adult pads set is InstalledPads set is installed CPR Coaching infant/child ventilation instructionTesting and troubleshooting TestingTroubleshooting Recommended action when you need to use the deviceTroubleshooting while the FRx is being used Defibrillator saysTo replace pads To make sure the pads have been removed from the caseDefibrillator says To stop all motion Troubleshooting while the FRx is not being used BehaviorAdditional technical data required for European conformity Electromagnetic conformityElectromagnetic emissions Emissions testElectromagnetic immunity Rated maximum output power of transmitter W 100 Separation distance according to frequency of transmitter mMHz to2.5 GHz = 1.15√ PAppendices Philips Medical Systems Intentionally blank Philips Medical Systems United States CanadaEurope, Middle East, and Africa Latin America

861304 specifications

The Philips 861304 is a versatile and advanced home appliance that blends functionality with sleek design, making it an ideal choice for modern households. This innovative product is particularly known for its impressive performance in the realm of kitchen technology.

One of the standout features of the Philips 861304 is its state-of-the-art cooking capabilities. Equipped with an advanced heating system, the appliance ensures even and precise cooking temperatures. The user can enjoy a variety of cooking modes, including steaming, baking, and grilling, which allow for diverse culinary experiments. This versatility is complemented by a robust set of presets that simplify meal preparation, making it accessible for even novice cooks.

In terms of technology, the Philips 861304 utilizes innovative Rapid Air Technology that circulates hot air around the food. This not only ensures even cooking but also enhances the flavor while reducing the need for excess oil, making meals healthier. The appliance also features an intuitive touch-screen interface, allowing users to select their desired cooking settings with ease. This user-friendly design is critical for those who are busy but still want to enjoy home-cooked meals without the stress of complicated controls.

Another key characteristic of the Philips 861304 is its commitment to safety. The appliance is equipped with multiple safety features, including an automatic shut-off function and cool-touch handles, ensuring peace of mind during operation. Its compact design makes it suitable for kitchens of all sizes, allowing users to enjoy high-end cooking technology without sacrificing counter space.

The Philips 861304 is also designed with cleanliness in mind. The non-stick surface ensures easy food release and simplifies the cleaning process. Many components are dishwasher-safe, further streamlining maintenance and promoting a hygienic cooking environment.

In summary, the Philips 861304 stands out in the crowded market of kitchen appliances due to its advanced heating technology, user-friendly interface, and commitment to safety and cleanliness. Its combination of versatility and performance makes it a valuable addition to any culinary space, empowering users to effortlessly create delicious and healthy meals.