Page 51

Philips Medical Systems

parameter

protocol pause timer (minutes)

NSA pause type

settings

0.5, 1.0, 1.5, 2.0, 2.5, 3.0

Standard NSA pause: FRx does not perform rhythm analysis during the NSA pause.

SMART NSA pause: FRx conducts background monitoring during the SMART NSA pause. If a potentially shockable rhythm is detected, FRx terminates the SMART NSA pause and resumes rhythm analysis.

43

default

default description

2.0A 2-minute protocol pause for CPR automatically starts after voice instruction is given when a shock series is completed.

After the protocol pause, the defibrillator returns to rhythm analysis.

If the user presses the i-button for optional CPR coaching, the FRx provides coaching for 5 cycles of CPR, starting and ending with compressions, when the CPR Coaching parameters are also set to their default values. The number of CPR cycles varies for other protocol pause timer and CPR Coaching parameter settings.

SMART NSA

During a SMART NSA pause,

pause

the defibrillator conducts

 

background monitoring. If a

 

potentially shockable rhythm

 

is detected in a motionless

 

patient, the defibrillator

 

terminates the SMART NSA

 

pause and resumes rhythm

 

analysis.

 

NOTE: If the FRx detects CPR in

 

progress or if the responder has

 

pressed the i-button for CPR

 

Coaching, the SMART NSA

 

pause will be converted to a

 

standard NSA pause. During the

 

standard NSA pause, the

 

defibrillator does not perform

 

rhythm analysis.

APPENDICES

Image 51
Contents 861304 N E R ’ S M a N U a LEdition Philips Medical Systems Quick Reference Intentionally blank HeartStart FRx Device Tracking About This Edition Edition HistoryCopyright Authorized EU RepresentativeContents Appendices Sudden Cardiac Arrest DescriptionIndications for Use Training and practiceState and local requirements For more information Package contents1A. Remove Pads Case from packaging Setting up the FRxPhilips Medical Systems Recommended accessories Overview Using the HeartStart FRxPress the green on/off button Follow the FRx’s voice instructionsOpen the pads case as shown below. Peel off one pad If a shock is needed Press the flashing orange Shock button if instructedIf a shock is not needed Treating infants and childrenKey if available Infant/Child Key When emergency medical services arrivePhilips Medical Systems After each use After using the HeartStart FRxFRx data storage Maintaining the HeartStart FRx Routine maintenanceCleaning the FRx Periodic checksDisposing of the FRx Troubleshooting tips Appendices Philips Medical Systems Accessories Appendices Glossary of terms Coaching, if selected, appropriate for infants and children HeartStart Event ReviewButton Infant/Child KeyPads See Smart Pads When a battery is insertedPatient care in the AED mode On/Off buttonWaveform See Smart biphasic waveform Standby modeBlinking green Ready light Consciousness, absence of respiration, and lack of a pulseGlossary of symbols/controls Standby mode gives device statusShock is advised, as a reminder not to touch the patient Symbol  #  Appendices Accessories Please read these warnings and precautions carefullyApproved accessories are used Hazard. Do not sterilize the FRx or its accessoriesPrecautions Environmental HeartStart FRx 861304 Defibrillator specificationsTechnical information PhysicalControls and indicators Category Specifications SealingProtected Current a Defibrillation waveformNewborn 14 J/kg Disarm AED mode Adult shock delivery vectorCategory Energy Charge control Shock-to-shock cycle time Energy doseECG analysis system ECG analysis performanceRhythm class Samplea size Performance Lower HeartStart Smart Pads II Accessories specificationsM5070A battery and 989803139301 TSO C-142* battery Category Battery type CapacityInfant/Child Key Environmental considerations EnvironmentRegulations Overview Device options ConfigurationPatient treatment protocol options Is detected in a motionless NSA pause type SettingsDefibrillator conducts Background monitoring. If aInstructs Instructions provided at User that it is CPR prompt SettingsCPR1 Instructs the user to begin CPR CPR reminder voiceCPR Coaching infant/child ventilation instruction When an adult pads set isInstalled Pads set is installedRecommended action when you need to use the device Testing and troubleshootingTesting TroubleshootingTo make sure the pads have been removed from the case Troubleshooting while the FRx is being usedDefibrillator says To replace padsDefibrillator says To stop all motion Behavior Troubleshooting while the FRx is not being usedEmissions test Additional technical data required for European conformityElectromagnetic conformity Electromagnetic emissionsElectromagnetic immunity = 1.15√ P Rated maximum output power of transmitter W 100Separation distance according to frequency of transmitter m MHz to2.5 GHzAppendices Philips Medical Systems Intentionally blank Latin America Philips Medical Systems United StatesCanada Europe, Middle East, and Africa

861304 specifications

The Philips 861304 is a versatile and advanced home appliance that blends functionality with sleek design, making it an ideal choice for modern households. This innovative product is particularly known for its impressive performance in the realm of kitchen technology.

One of the standout features of the Philips 861304 is its state-of-the-art cooking capabilities. Equipped with an advanced heating system, the appliance ensures even and precise cooking temperatures. The user can enjoy a variety of cooking modes, including steaming, baking, and grilling, which allow for diverse culinary experiments. This versatility is complemented by a robust set of presets that simplify meal preparation, making it accessible for even novice cooks.

In terms of technology, the Philips 861304 utilizes innovative Rapid Air Technology that circulates hot air around the food. This not only ensures even cooking but also enhances the flavor while reducing the need for excess oil, making meals healthier. The appliance also features an intuitive touch-screen interface, allowing users to select their desired cooking settings with ease. This user-friendly design is critical for those who are busy but still want to enjoy home-cooked meals without the stress of complicated controls.

Another key characteristic of the Philips 861304 is its commitment to safety. The appliance is equipped with multiple safety features, including an automatic shut-off function and cool-touch handles, ensuring peace of mind during operation. Its compact design makes it suitable for kitchens of all sizes, allowing users to enjoy high-end cooking technology without sacrificing counter space.

The Philips 861304 is also designed with cleanliness in mind. The non-stick surface ensures easy food release and simplifies the cleaning process. Many components are dishwasher-safe, further streamlining maintenance and promoting a hygienic cooking environment.

In summary, the Philips 861304 stands out in the crowded market of kitchen appliances due to its advanced heating technology, user-friendly interface, and commitment to safety and cleanliness. Its combination of versatility and performance makes it a valuable addition to any culinary space, empowering users to effortlessly create delicious and healthy meals.