Calibration procedure by sensor type
i) Hi-Sens Sensors
1)Ensure that the sensor is clean and dry, that nothing is touching the sensor body, and that the delay adjustment on the control unit is fully
2)Set the gain potentiometer in the head amplifier fully
3)The control unit should now indicate ‘alarm’ (or ‘normal’ for inverse acting head amplifiers).
4)Gradually increase the gain until the control unit indicates ‘normal’ (or ‘alarm’ for inverse acting control units). This is beginning of the operating band.
5)The optimum gain setting can be found by rotating the potentiometer a further 30º (one division on the clock face),
6)Check that the control unit switches satisfactorily when the sensor is immersed in the liquid.
ii) Gap Sensors
1)Ensure that the sensor is clean, and immerse it in a sample of the liquid, making sure that the gap is full of the liquid, and unobstructed. Adjust the time delay on the control unit fully anti- clockwise.
2)Set the head amplifier gain control fully
3)The control unit should now indicate alarm (or normal for inverse acting head amplifiers) if it does not switch to low gain (if a gain switch is fitted).
4)Gradually increase the gain until the control unit indicates normal (or alarm for inverse acting head amplifiers). Note the position of the potentiometer and gain switch. This is the lower end of the working range.
5)Remove the sensor from the liquid, and allow the liquid to drain away. Ensure that nothing is touching the sensor body. The control unit should now indicate alarm (or normal for inverse acting head amplifiers).
6)Gradually increase the gain until the control unit indicates normal (or alarm for inverse acting head amplifiers). Note the position of the potentiometer and gain switch. This is the upper limit of the working range.
7)The optimum gain setting is midway between the points recorded in steps 4 and 6. It may be necessary to estimate this point if the gain switch has been used.
8)Check that the control unit switches correctly when the sensor is immersed in the liquid.
iii) Interface Sensors – Attenuation Method
1)Ensure that the sensor is clean, and that the delay on the control unit is fully
2)With the sensor in the lower liquid adjust the gain potentiometer
3)Note the position of the potentiometer. If the control unit does not switch to ‘Alarm’ use the bottom end of the scale as a reference.
4)Immerse the sensor in the upper liquid. The control unit should now indicate ‘Alarm’. If it does not then the reflection method should be used.
8