16
Sensor Alignment
RH Sensor Alignment
6.Place the Salt container, with the salt alignment cap removed, over the RH sensor (middle sensor). Allow level to stabilize for at least 15 minutes. (Refer to Figure
∙About sensor alignment: if readings are within
NOTE: For a High alignment, use the 75% sensor alignment salt (NaCL) and for a Low alignment, use the 33% alignment salt (MgCl).
Alignment salt (remove rubber cap and place over RH sensor with the sunshield removed)
Once Alignment salt is placed, allow levels to stabilize (about 15 mins).
Low RH Alignment screen Adjust % to 33% (using up/down arrow
keys) and press Enter (after level has stabilized) to store sensor alignment level.
Figure 1-27: RH alignment with Low % example
7.In either the High/Low RH alignment screen, adjust the level to the percentage displayed on the salt container by using the Up/Down Arrow keys. Press I/O Enter key to save.
(The instrument will state either “Successful” or “Failure” if it passed/did not pass. For
“Failure” repeat the steps above.)
Press Enter to save the alignment.
Example of high RH percentage prior to sensor alignment. Press up/down arrows to adjust level after salt has stabilized.
Sensor alignment with High % RH level
Figure 1-28: RH alignment with High% screen
8.To return to the main menu, press Run/Stop key.
9.Verify the RH sensor is
10.Remove the alignment salt and finger cot. Replace the Sunshield over the RH sensor when completed with the sensor alignment.