Teledyne 760 PLUS instruction manual Pressure Measurements, Auxiliary Relay Contacts

Page 11

Auxiliary Relay Contacts

The THPS-760 Power Supply has SP1 and SP2 relays for various external switching/monitoring applications. Each relay has a single set of Form “C” contacts. (Choice of normally open or normally closed contacts). The relays will energize (pick up) when the input is less than the preset set points as adjusted in the Set Point Adjustment procedures above. The set point LED Indicators (SP1 or SP2) will illuminate indicating that (SP1 or SP2) relay is energized. The choice of the normally open or normally closed contacts will depend on the desired switching function of the particular application. Each set of contacts is capable of switching up to 2 amps (max) @ 250 vac .

3.7. Pressure Measurements

The 760 Plus piezo transducers are shipped from the factory fully calibrated; thus, can provide immediate and accurate pressure measurements as soon as the installation outlined in Section 2.2 has been completed. If using a 760 Plus power supply/display, the pressure can be read directly in Torr units. When not using a 760 Plus power supply/display, the linear output voltage or current depending on option, must be interpolated to obtain absolute pressure measurements.

Figure 3-3 on the following page of this manual shows a typical signal output vs pressure curve. Note the excellent linearity over 4-decades of pressure.

The Hastings 760 Plus power supply/display is calibrated at the factory prior to shipment and does not require adjustment when installed. The user should not attempt to calibrate the Hastings 760 Plus power supply/display; however, if the display indication is incorrect, calibrate the Hastings piezo transducer by adjusting the ZERO and SPAN controls as outlined in this manual.

The Hastings transducer can be adjusted by performing the following procedure. If you intend to use this procedure, you will need a vacuum system capable of achieving base pressures below 1.00E-02 Torr, and a power supply/display which is calibrated for the option in use.

1000.0

 

 

 

5.0

100.0

 

 

 

0.5

INDICATED

 

 

 

OUTPUT

PRESSURE

 

 

 

 

 

 

VOLTS

(TORR)

 

 

 

 

 

 

 

10.0

 

 

 

0.05

1.0

 

 

 

0.005

0.1

 

 

 

0.0005

0.10

1.00

10.00

100.00

1000.00

PRESSURE (TORR)

Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge

Page 11 of 16

Image 11
Contents Teledyne Hastings Manual Print History Table of Contents Piezo Vacuum Transducer 760s FeaturesPower Supply/Display General InformationOutput signal options SpecificationsReceiving Inspection InstallationPiezo Transducer 760s Installation Plus Interconnect Cable Power Supply/Display InstallationPower Supply/Display Cables and Wiring General Operating InformationTHPS-760 Power Supply Physical Layout Piezo Transducer Input PowerPiezo Transducer Output Signal Power Supply/Display Input PowerGrounding Considerations Power Supply Relay Set Point AdjustmentJ3 Input/Output Connector Auxiliary Relay Contacts Pressure MeasurementsCalibration Hastings 760 Plus Piezo Transducer Zero and Span AdjustmentHastings 760s Option-5 Transducer Zero and Span Adjustment Piezo Transducer Authorized Maintenance MaintenancePower Supply/Display Authorized Maintenance Non-Warranty Repair Policy Warranty Repair PolicyRear Panel Diagrams and DrawingsFront Panel Panel Mounting Mounting