Teledyne 760 PLUS instruction manual Piezo Transducer Input Power, Piezo Transducer Output Signal

Page 9

3.2. Piezo Transducer Input Power

The 760s piezo transducer operates with a DC input from 15 to 30 volts. Use the 4-pin female connector supplied and makes necessary connections as outlined. See illustration in Section 2 of this manual.

3.3. Piezo Transducer Output Signal

The 760s piezo transducer provides either a linear DC voltage output with pressure or a linear DC current output with pressure. The type of output is determined by the instrument option specified. A brief description of the option, pressure range and output signal is listed below:

OPTION No

PRESSURE RANGE

SIGNAL OUT

 

(Torr)

(see note

01

0-1000

0-5 VDC

02

0-1000

0-10 VDC

03

0-1000

2-10 mA

04

0-1000

4-20 mA

05

0-1000

0-5 or 10 VDC

w/LCD Digital Display

NOTE: The 760 Plus power supply/ display or the 760s option-05, provide a direct pressure readout in Torr units.

3.4. Power Supply/Display Input Power

The 760 Plus power supply/display must be connected to a source of AC voltage, either 120 or 240 V, before attempting to program or setup the unit for operation with the Hastings 760S piezo vacuum transducer (see Section 2.3 for cable description).

3.5.THPS-760 Power Supply Physical Layout

3.5.1.Front Panel (Refer to page 19, 7.1.1 for physical layout.)

Digital Display. The THPS-760 has a 4 1/2-digit display that displays continuous readout.

Set Point Adjustment Pots.

Two Set Point Adjustment Pots are accessible on the panel and are used to set the desired set points. (SP1/SP2) Refer to SET POINT ADJUSTMENTS and figure 1 for specific instructions pertaining to setting the set points.

THE KNOBS ON THE SET POINT ADJUSTMENT POTS ARE PERMANENTLY INSTALLED AND CAN NOT BE REMOVED. REMOVING THEM COULD CAUSE PERMANENT DAMAGE TO THE SET POINT POTS.

Set Point Pushbuttons.

There are two set point pushbuttons (SP1 AND SP2) accessible on the front panel. When depressed, the setpoint value is displayed on the LED readout

SET POINT LED’s. The two LED indicators (SP1 and SP2) will illuminate when the input pressure is less than desired (SP1/SP2) set point. The LED’s indicate the corresponding relay is activated (energized).

Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge

Page 9 of 16

Image 9
Contents Teledyne Hastings Manual Print History Table of Contents Power Supply/Display FeaturesGeneral Information Piezo Vacuum Transducer 760sOutput signal options SpecificationsInstallation Piezo Transducer 760s InstallationReceiving Inspection Power Supply/Display Installation Power Supply/Display Cables and WiringPlus Interconnect Cable General Operating InformationPiezo Transducer Output Signal Piezo Transducer Input PowerPower Supply/Display Input Power THPS-760 Power Supply Physical LayoutPower Supply Relay Set Point Adjustment J3 Input/Output ConnectorGrounding Considerations Auxiliary Relay Contacts Pressure MeasurementsHastings 760 Plus Piezo Transducer Zero and Span Adjustment Hastings 760s Option-5 Transducer Zero and Span AdjustmentCalibration Maintenance Power Supply/Display Authorized MaintenancePiezo Transducer Authorized Maintenance Non-Warranty Repair Policy Warranty Repair PolicyDiagrams and Drawings Front PanelRear Panel Panel Mounting Mounting