Teledyne 760 PLUS instruction manual Receiving Inspection, Piezo Transducer 760s Installation

Page 6

2. Installation

This section contains the necessary instructions to assist in placing the HPM-760 Plus digital vacuum gauge into operation as quickly and easily as possible. Please read the following thoroughly before attempting to install or use the instrument.

2.1. Receiving Inspection

Carefully unpack the Hastings product(s). Inspect for any obvious signs of damage to the shipment. Immediately advise the carrier who delivered the shipment if any damage is suspected. Check each component shipped with the packing list. Insure that all parts listed are present. Other equipment or accessories will be listed separately.

2.2. Piezo Transducer (760s) Installation

Figure 2-1 shows the Hastings piezo vacuum transducer. A standard 1/8” NPT male stem and

evacuation port is provided for system connection. VCR and KF stem terminations are also available

When installing transducers with threaded stem, apply an approved vacuum sealant to the stem threads of the transducer. Hand thread transducer into vacuum system until joint is finger tight. Tighten transducer module completely using a wrench on the hexagonal portion of the stem.

Do not grip instrument case to tighten; this will result in permanent damage to transducer assembly. When the 760s is purchased independently of the power supply/display, a connector is provided so that the transducer can be connected as detailed in the figure below.

TRANSDUCER PINOUT

 

PIN1

SIGNAL OUTPUT (COM)

PIN 2

SIGNAL OUTPUT (POS)

PIN 3

15-30 VDC INPUT (COM)

PIN 4

15-30 VDC INPUT (POS)

Manual: 147-082010_HPM-760-Plus-Vacuum-Gauge

Page 6 of 16

Image 6
Contents Teledyne Hastings Manual Print History Table of Contents General Information FeaturesPower Supply/Display Piezo Vacuum Transducer 760sSpecifications Output signal optionsInstallation Piezo Transducer 760s InstallationReceiving Inspection Power Supply/Display Installation Power Supply/Display Cables and WiringPlus Interconnect Cable Operating Information GeneralPower Supply/Display Input Power Piezo Transducer Input PowerPiezo Transducer Output Signal THPS-760 Power Supply Physical LayoutPower Supply Relay Set Point Adjustment J3 Input/Output ConnectorGrounding Considerations Pressure Measurements Auxiliary Relay ContactsHastings 760 Plus Piezo Transducer Zero and Span Adjustment Hastings 760s Option-5 Transducer Zero and Span AdjustmentCalibration Maintenance Power Supply/Display Authorized MaintenancePiezo Transducer Authorized Maintenance Warranty Repair Policy Non-Warranty Repair PolicyDiagrams and Drawings Front PanelRear Panel Mounting Panel Mounting