4.Calibration
4.1.Hastings 760 Plus Piezo Transducer Zero and Span Adjustment
1.Install the transducer and power supply/display then apply operating power. Wait a mini mum of 15 minutes before proceeding.
2.Evacuate system to a pressure less than 0.010 Torr
3.Adjust transducer zero trimpot (Fig.
4.Increase system pressure to atmosphere and adjust span trimpot until the power supply/ display indicates the known local atmospheric pressure ±000.1 Torr.
5.Repeat steps 2 through 4 if necessary.
4.2.Hastings 760s Option-5 Transducer Zero and Span Adjustment
1.Install the transducer and apply operating power. Wait a minimum of 15 minutes before proceeding.
2.Evacuate the system to a pressure less than 0.01 Torr
3.Adjust transducer zero trimpot (Fig.
4.Increase system pressure to atmosphere and adjust span trimpot (Fig.
5.Repeat steps 2 through 4 if necessary.