Braun HNS 11 user manual LCD display, Pulse LED display

Page 10

10

Pulse LED display

The small orange LED above the LCD display flashes according to the pulse frequency set (1 Hz: once per second; 2 Hz: twice per second) when current is flowing through the electrodes. At the same time, a louder tick can be heard.

The LCD display

According to the display mode selected, the LCD display shows the setting for, or the actually flowing stimulation current. The display is in mA.

The ▲ symbol indicates the stimulation frequence. At the same time, there is a ticking sound in the same stimulation frequence. A ▲ symbol appears next to the ~ symbol in the same stimulation frequence when the stimulation current is flowing and the ticking sound gets louder. When the actual value display is selected, the LCD display additionally shows a "+" in front of the current displayed.

Exceptions:

1.Pressing the ”BAT” button displays the battery voltage.

2.Equipment-related error functions produce the corresponding error code in the display (see p. 17).

3.The ”BAT” symbol in the display indicates that the battery must be exchanged soon.

4.When the battery voltage drops to approx. 6 volts

or

appears on the display. The stimulator switches off automatically when these signs appear.

Display:

The ”BAT” symbol indicates that the battery voltage is too low. (Refer to the ”Battery” section, p. 13)

Image 10
Contents Stimuplex HNS Manufacturer Distributor Stockert GmbH Contents Checks before start-up Page Technical description Selecting the pulse frequency Switching on / switching off Operating the nerve stimulator Selecting the pulse widthSelecting the current display mode Checking the battery voltageSelecting the current range LED displays Current display modeCurrent adjustment range Pulse widthLCD display Pulse LED displayTesting the Stimuplex HNS 11 before an application Initial set up, maintenance, and general adviceTone Special technical features Technical dataBattery Changing the batteryCleaning and disinfecting the Stimuplex HNS Maintenance and safety checksPage Warranty Safety information Error messagesStimuplex D needles, 15 bevel Instruments and accessoriesProduct discription Canula Code number Sales unit Length Stimuplex D needles 30 bevelContiplex D needles, 15 bevel Product discription Canula Code numberLength Stimuplex a needles, 30 bevel Contiplex D needles, 30 bevelPeripheral electrical nerve stimulation PrinciplesPage Page Literature Appendix a Appendix BAppendix C Appendix D Only 0.2 or 0.05 mA Stimulation current requiredPage Hospital Care B. Braun Melsungen AG