14
Cleaning and disinfecting the Stimuplex® HNS 11
Use only soft, damp cloths to clean and disinfect the Stimuplex® HNS 11 and the electrode cable. Water, soap or white spirits are especially suitable for this. Ensure that no moisture penetrates the Stimuplex® HNS 11.
Use only wet disinfection, no spray disinfection! Avoid condensation!
White spirit or commercially available
Caution!
The following substances must not be used:
trichloroethylene, acetone, methyl ethyl ketone, benzene, metha- nol, cellulose thinner,
Maintenance and safety checks
Check the serviceable condition of the Stimuplex® HNS 11 and ac- cessories on each occasion before use. A defective instrument must not be used.
Electromedical equipment may only be repaired by the manufac- turer or by an organization expressly authorized by the manufac- turer. An order for repairs must be accompanied by a detailed description of the fault (acc. separate Technical Service Manual).