Appendix A | 24 |
|
|
Stimulation current display in the mA range
Patient current | Measurement display | Controller display |
| Patient current | Measurement display | Controller display | ||||
0.00 mA | 0.00 mA |
| 0.00 mA |
|
| 0.00 mA | 0.00 mA |
| 0.00 mA |
|
0.05 mA | 0.05 mA ± | 0.02 | 0.05 mA ± | 0.03 |
| 0.50 mA | 0.50 mA ± | 0.1 | 0.50 mA ± | 0.15 |
0.10 mA | 0.10 mA ± | 0.02 | 0.10 mA ± | 0.03 |
| 1.00 mA | 1.00 mA ± | 0.1 | 1.00 mA ± | 0.15 |
0.20 mA | 0.20 mA ± | 0.02 | 0.20 mA ± | 0.03 |
| 1.50 mA | 1.50 mA ± | 0.1 | 1.50 mA ± | 0.15 |
0.30 mA | 0.30 mA ± | 0.02 | 0.30 mA ± | 0.03 |
| 2.00 mA | 2.00 mA ± | 0.1 | 2.00 mA ± | 0.15 |
0.40 mA | 0.40 mA ± | 0.02 | 0.40 mA ± | 0.03 |
| 3.00 mA | 3.00 mA ± | 0.1 | 3.00 mA ± | 0.15 |
0.60 mA | 0.60 mA ± | 0.02 | 0.60 mA ± | 0.03 |
| 4.00 mA | 4.00 mA ± | 0.1 | 4.00 mA ± | 0.15 |
0.80 mA | 0.80 mA ± | 0.02 | 0.80 mA ± | 0.03 |
| 5.00 mA | 5.00 mA ± | 0.1 | 5.00 mA ± | 0.15 |
1.00 mA | 1.00 mA ± | 0.02 | 1.00 mA ± | 0.03 |
|
|
|
|
|
|
Appendix B
Conversion table: Amplitude (mA) and pulse width (ms) to charge
Amplitude | Pulse width | Charge | Pulse width | Charge | Pulse width | Charge |
mA | ms | nC | ms | nC | ms | nC |
0.1 | 0.1 | 10 | 0.3 | 30 | 1.0 | 100 |
0.2 | 0.1 | 20 | 0.3 | 60 | 1.0 | 200 |
0.3 | 0.1 | 30 | 0.3 | 90 | 1.0 | 300 |
0.4 | 0.1 | 40 | 0.3 | 120 | 1.0 | 400 |
0.5 | 0.1 | 50 | 0.3 | 150 | 1.0 | 500 |
0.6 | 0.1 | 60 | 0.3 | 180 | 1.0 | 600 |
0.7 | 0.1 | 70 | 0.3 | 210 | 1.0 | 700 |
0.8 | 0.1 | 80 | 0.3 | 240 | 1.0 | 800 |
0.9 | 0.1 | 90 | 0.3 | 270 | 1.0 | 900 |
1.0 | 0.1 | 100 | 0.3 | 300 | 1.0 | 1000 |
1.2 | 0.1 | 120 | 0.3 | 360 | 1.0 | 1200 |
1.6 | 0.1 | 160 | 0.3 | 480 | 1.0 | 1600 |
1.8 | 0.1 | 180 | 0.3 | 540 | 1.0 | 1800 |
2.0 | 0.1 | 200 | 0.3 | 600 | 1.0 | 2000 |
2.5 | 0.1 | 250 | 0.3 | 750 | 1.0 | 2500 |
3.0 | 0.1 | 300 | 0.3 | 900 | 1.0 | 3000 |
3.5 | 0.1 | 350 | 0.3 | 1050 | 1.0 | 3500 |
4.0 | 0.1 | 400 | 0.3 | 1200 | 1.0 | 4000 |
4.5 | 0.1 | 450 | 0.3 | 1350 | 1.0 | 4500 |
5.0 | 0.1 | 500 | 0.3 | 1500 | 1.0 | 5000 |