Page 13

13

Battery

The battery must be regularly checked for correct charge by press- ing the ”BAT” button. While the button is pressed, the existing battery voltage appears on the LCD display. The battery should be replaced under 7 volts – as soon as the BAT symbol appears.

When the battery voltage drops below 6 V, the nerve stimulator stops operating. Four horizontal lines and the BAT symbol appear in the LCD display. Change the battery.

When the Stimuplex® HNS 11 nerve stimulator is not used for a long period of time, the battery must be removed to prevent leak- age.

Changing the battery

The battery compartment is in the base of the nerve stimulator. Open the compartment by lightly pressing the OPEN symbol. At the same time, slide the cover in the direction of the arrow.

Change the battery, ensuring correct polarity. After fitting the bat- tery, slide the cover of the battery compartment back on again right up to the limit stop.

Use only 9 V alkaline manganese batteries (e.g. VARTA 4022, DURACELL MN 1604). These will provide you with a greatly extended operating time and thus considerably more reliable function.

Warning!

If the battery is leaking, the Stimuplex® HNS 11 should no longer be operated for safety reasons. Acid penetrating the inside of the instrument can damage or impair essential circuitry. It must be returned to the manufacturer for correct cleaning and a safety check.

Image 13
Contents Stimuplex HNS Manufacturer Distributor Stockert GmbH Contents Checks before start-up Page Technical description Operating the nerve stimulator Switching on / switching offSelecting the pulse frequency Selecting the pulse widthSelecting the current display mode Checking the battery voltageSelecting the current range Current adjustment range Current display modeLED displays Pulse widthPulse LED display LCD displayTesting the Stimuplex HNS 11 before an application Initial set up, maintenance, and general adviceTone Technical data Special technical featuresChanging the battery BatteryMaintenance and safety checks Cleaning and disinfecting the Stimuplex HNSPage Warranty Error messages Safety informationProduct discription Canula Code number Sales unit Length Instruments and accessoriesStimuplex D needles, 15 bevel Stimuplex D needles 30 bevelLength Stimuplex a needles, 30 bevel Product discription Canula Code numberContiplex D needles, 15 bevel Contiplex D needles, 30 bevelPrinciples Peripheral electrical nerve stimulationPage Page Literature Appendix B Appendix aAppendix C Only 0.2 or 0.05 mA Stimulation current required Appendix DPage Hospital Care B. Braun Melsungen AG